Development of a New Low-Energy Neutron Spectrometer Using a Position-Sensitive Proportional Counter: Examination of the Response Function Considering the Spectrometer Shield

Background: Boron neutron capture therapy (BNCT) is a potential radiation therapy for treating refractory cancers. Measuring the neutron energy spectrum is important to evaluate the biological effect of BNCT. We have developed a novel low-energy neutron spectrometer with a 3He position-sensitive proportional counter. The effectiveness of this spectrometer remained unvalidated due to the inadequate response function.Materials and Methods: The response function of this spectrometer was evaluated by Monte Carlo simulation to perform the validation experiment of this spectrometer. The 3He(n,p)3H reaction rate for each monochromatic neutron (energy range, 0.5 eV–10 keV) was calculated. Numerical experiments were conducted to investigate the performance of this spectrometer. Several estimated values of the energy spectrum were obtained and compared with a calculated energy spectrum at the exit of the epithermal column by the Monte Carlo simulation by preparing several candidates for signals. The Bayesian estimation algorithm was utilized for the estimation.Results and Discussion: The results indicate that the spectrometer and the evaluated response functions can be utilized to measure the energy spectrum of neutrons in the low-energy range. However, the energy spectrum could not be well represented >4 keV due to the poor response function, including uncertainties in response function calculation as well as the poor energy resolution of this detector in the high-energy region.Conclusion: The response function of the low-energy neutron spectrometer for the epithermal energy region can be assessed by the Monte Carlo simulation. The numerical experiments reveal that the effectiveness of this spectrometer can be shown in the range of 0.5 eV.

Authors

Institutions

Publication Details

Journal
Journal of Radiation Protection and Research
Published
2025-02-04
DOI
https://doi.org/10.14407/jrpr.2023.00325
Citations
1
Primary Topic
Nuclear Physics and Applications
Type
article
Field-Weighted Citation Impact
1.88
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a New Low-Energy Neutron Spectrometer Using a Position-Sensitive Proportional Counter: Examination of the Response Function Considering the Spectrometer Shield

Isao Murata, Yu Fujiwara, Fuminobu Sato, Shingo Tamaki
1 citations
Journal of Radiation Protection and Research
Nuclear Physics and Applications
1.88
article

Development of a New Low-Energy Neutron Spectrometer Using a Position-Sensitive Proportional Counter: Examination of the Response Function Considering the Spectrometer Shield

Isao Murata, Yu Fujiwara, Fuminobu Sato, Shingo Tamaki
article en
1 citations

Abstract

Background: Boron neutron capture therapy (BNCT) is a potential radiation therapy for treating refractory cancers. Measuring the neutron energy spectrum is important to evaluate the biological effect of BNCT. We have developed a novel low-energy neutron spectrometer with a 3He position-sensitive proportional counter. The effectiveness of this spectrometer remained unvalidated due to the inadequate response function.Materials and Methods: The response function of this spectrometer was evaluated by Monte Carlo simulation to perform the validation experiment of this spectrometer. The 3He(n,p)3H reaction rate for each monochromatic neutron (energy range, 0.5 eV–10 keV) was calculated. Numerical experiments were conducted to investigate the performance of this spectrometer. Several estimated values of the energy spectrum were obtained and compared with a calculated energy spectrum at the exit of the epithermal column by the Monte Carlo simulation by preparing several candidates for signals. The Bayesian estimation algorithm was utilized for the estimation.Results and Discussion: The results indicate that the spectrometer and the evaluated response functions can be utilized to measure the energy spectrum of neutrons in the low-energy range. However, the energy spectrum could not be well represented >4 keV due to the poor response function, including uncertainties in response function calculation as well as the poor energy resolution of this detector in the high-energy region.Conclusion: The response function of the low-energy neutron spectrometer for the epithermal energy region can be assessed by the Monte Carlo simulation. The numerical experiments reveal that the effectiveness of this spectrometer can be shown in the range of 0.5 eV.

Journal of Radiation Protection and ResearchVol. 50(s1)
The University of Osaka (JP)
Affordable and clean energy
Openalex Percentile: Top 21%
Nuclear Physics and Applications
1.88
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.