3D-printed boluses in radiation therapy are of increasing interest for enhancing treatment precision and patient comfort. A comprehensive clinical validation of these boluses remains to be established. This study aims to confirm the effectiveness of a 3D-printed bolus through a proof-of-concept comparative validation, by implementing in a clinical setting a bolus made of PLA and designed to ensure uniform dose coverage for a case in the eye region. In this study the 3D-printed bolus was compared to two commercially available boluses (one thermoplastic and one skin type) by using a refecence where no bolus was present (with the optimal dose distribution scenario). All boluses were placed on an anthropomorphic head phantom and BeOSL detectors were used to measure dose values to determine the level of their effectiveness on delivery. During the scanning process, a thermoplastic mask was used to prevent bolus movement and to accurately reproduce clinical scenarios. Differences in dose values at D<sub>max</sub>and D<sub>50%</sub>revealed the performance of each bolus. The treatment planning system (TPS) and BeOSL readings for the 3D printed bolus were within 2% (the clinical tolerance), with 0.66% dose difference for the customized 3D-printed bolus. Although the thermoplastic bolus had the closest value to the detector reading, with a score of 0.30%, this result was influenced by improper shaping of the bolus on the phantom and the presence of a wide air gap, which caused lack of eye covering. Whereas, the skin bolus, due to higher volume of air between phantom surface and bolus, showed a 1.29% dose difference between the TPS values and the OSL detector readings. We provide a comparative validation for the use of 3D printed boluses and highlight that proper bolus fitting is essential in clinical settings to avoid air gaps and to maintain dose distribution over multiple treatment sessions.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2025-02-03
DOI
https://doi.org/10.1088/2057-1976/adb15d
Citations
2
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
3.77
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation of a 3D printed bolus for radiotherapy: a proof-of-concept study

Lucian Cristian Petcu, Ferenc Járai-Szabó, A. Ciobanu, Zoltán Bálint
2 citations
Biomedical Physics & Engineering Express
Advanced Radiotherapy Techniques
3.77
article

Validation of a 3D printed bolus for radiotherapy: a proof-of-concept study

Lucian Cristian Petcu, Ferenc Járai-Szabó, A. Ciobanu, Zoltán Bálint
article en
2 citations

Abstract

3D-printed boluses in radiation therapy are of increasing interest for enhancing treatment precision and patient comfort. A comprehensive clinical validation of these boluses remains to be established. This study aims to confirm the effectiveness of a 3D-printed bolus through a proof-of-concept comparative validation, by implementing in a clinical setting a bolus made of PLA and designed to ensure uniform dose coverage for a case in the eye region. In this study the 3D-printed bolus was compared to two commercially available boluses (one thermoplastic and one skin type) by using a refecence where no bolus was present (with the optimal dose distribution scenario). All boluses were placed on an anthropomorphic head phantom and BeOSL detectors were used to measure dose values to determine the level of their effectiveness on delivery. During the scanning process, a thermoplastic mask was used to prevent bolus movement and to accurately reproduce clinical scenarios. Differences in dose values at D<sub>max</sub>and D<sub>50%</sub>revealed the performance of each bolus. The treatment planning system (TPS) and BeOSL readings for the 3D printed bolus were within 2% (the clinical tolerance), with 0.66% dose difference for the customized 3D-printed bolus. Although the thermoplastic bolus had the closest value to the detector reading, with a score of 0.30%, this result was influenced by improper shaping of the bolus on the phantom and the presence of a wide air gap, which caused lack of eye covering. Whereas, the skin bolus, due to higher volume of air between phantom surface and bolus, showed a 1.29% dose difference between the TPS values and the OSL detector readings. We provide a comparative validation for the use of 3D printed boluses and highlight that proper bolus fitting is essential in clinical settings to avoid air gaps and to maintain dose distribution over multiple treatment sessions.

Biomedical Physics & Engineering ExpressVol. 11(2)
Ovidius University (RO), Babeș-Bolyai University (RO)
Good health and well-being
Openalex Percentile: Top 10%
Advanced Radiotherapy Techniques
3.77
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.