Study of attenuation characteristics for novel neonatal head phantom in diagnostic radiology using Monte Carlo simulations and experiments

This study presents the design and validation of a neonatal head phantom using innovative heterogeneous composite materials customized to replicate the x-ray attenuation properties of neonatal cranial structures. Analysis of Hounsfield Unit (HU) data from 338 neonatal head CT scans informed the design of epoxy resin-based composites with additives such as sodium bicarbonate, fumed silica, and acetone to simulate bone, brain matter, cerebrospinal fluid (CSF) and hyperdense abnormalities. The cranial bone substitute (60% epoxy resin, 40% sodium bicarbonate) achieved a density of 1.60 g cm<sup>-3</sup>, with HU values (574.67-608.04) closely matching clinical ranges. Brain matter (95% epoxy resin, 5% acetone) achieved HU values (35.27-43.61), aligning with clinical means, while the CSF-equivalent material (80% epoxy resin, 15% fumed silica, 5% acetone) matched neonatal CSF HU values (14.53-17.02). A mass substitute for hyperdense abnormalities exhibited HU values (56.16-61.07), enabling differentiation from normal brain. Validation included Monte Carlo simulations and experimental CT imaging, showing close agreement in linear attenuation coefficients, with deviations below 11% across energy levels. Mass attenuation coefficients from simulations and XCOM software were consistent, with deviations under 0.7%, confirming the materials dosimetric reliability. The phantom, with a cylindrical geometry (9 cm diameter, 10 cm length), provides accurate attenuation properties across 80-120 kVp energy levels, with deviations below 5% between experimental CT numbers and simulation data. This phantom offers a robust platform for neonatal imaging research, enabling impactful dose optimization and imaging protocol adjustment and supports improved diagnostic accuracy in pediatric imaging.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2025-02-03
DOI
https://doi.org/10.1088/2057-1976/adb15c
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Study of attenuation characteristics for novel neonatal head phantom in diagnostic radiology using Monte Carlo simulations and experiments

H. Sekkat, A. Khallouqi, Youssef Madkouri, Omar El Rhazouani et al.
Biomedical Physics & Engineering Express
Medical Imaging Techniques and Applications
article

Study of attenuation characteristics for novel neonatal head phantom in diagnostic radiology using Monte Carlo simulations and experiments

H. Sekkat, A. Khallouqi, Youssef Madkouri, Omar El Rhazouani, Abdellah Halimi
article en

Abstract

This study presents the design and validation of a neonatal head phantom using innovative heterogeneous composite materials customized to replicate the x-ray attenuation properties of neonatal cranial structures. Analysis of Hounsfield Unit (HU) data from 338 neonatal head CT scans informed the design of epoxy resin-based composites with additives such as sodium bicarbonate, fumed silica, and acetone to simulate bone, brain matter, cerebrospinal fluid (CSF) and hyperdense abnormalities. The cranial bone substitute (60% epoxy resin, 40% sodium bicarbonate) achieved a density of 1.60 g cm<sup>-3</sup>, with HU values (574.67-608.04) closely matching clinical ranges. Brain matter (95% epoxy resin, 5% acetone) achieved HU values (35.27-43.61), aligning with clinical means, while the CSF-equivalent material (80% epoxy resin, 15% fumed silica, 5% acetone) matched neonatal CSF HU values (14.53-17.02). A mass substitute for hyperdense abnormalities exhibited HU values (56.16-61.07), enabling differentiation from normal brain. Validation included Monte Carlo simulations and experimental CT imaging, showing close agreement in linear attenuation coefficients, with deviations below 11% across energy levels. Mass attenuation coefficients from simulations and XCOM software were consistent, with deviations under 0.7%, confirming the materials dosimetric reliability. The phantom, with a cylindrical geometry (9 cm diameter, 10 cm length), provides accurate attenuation properties across 80-120 kVp energy levels, with deviations below 5% between experimental CT numbers and simulation data. This phantom offers a robust platform for neonatal imaging research, enabling impactful dose optimization and imaging protocol adjustment and supports improved diagnostic accuracy in pediatric imaging.

Biomedical Physics & Engineering ExpressVol. 11(2)
Université Ibn-Tofail (MA), Université Hassan 1er (MA), Hassania School of Public Works (MA)
Affordable and clean energy
Openalex Percentile: Top 97%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.