Effects of key parameters on pulsed field ablation of atrial fibrillation: potato experiments
The key ablation parameters have varying degrees of influence on lesion depths, following the order of PW>V; V>P; PW>P; N>P. Lesion depths increased with higher values of V and P. However, when the total number of pulses reached 9,600, there was no significant increase in injury depth.
Authors
- Yunhao Li (ORCID: https://orcid.org/0000-0002-5359-2930)
- Ping Zhang (ORCID: https://orcid.org/0000-0002-2295-9170)
- Qi Zhang (ORCID: https://orcid.org/0000-0002-8935-8396)
- Yaling Han
- Mingyu Sun (ORCID: https://orcid.org/0000-0001-8476-7046)
- Zulu Wang (ORCID: https://orcid.org/0009-0006-0364-8736)
- Daoyang Zhang
- Ming Liang
- Jie Zhang (ORCID: https://orcid.org/0009-0004-9779-1810)
- Fengqi Xuan
Institutions
- Dalian Medical University (CN)
- General Hospital of Shenyang Military Region (CN)
- Tianjin Chest Hospital (CN)
- China Medical University (CN)
Publication Details
- Journal
- Frontiers in Cardiovascular Medicine
- Published
- 2025-02-04
- DOI
- https://doi.org/10.3389/fcvm.2025.1525762
- Citations
- 4
- Primary Topic
- Microbial Inactivation Methods
- Type
- article
- Field-Weighted Citation Impact
- 4.93