Phantom-based investigation of block sequential regularised expectation maximisation (BSREM) reconstruction for zirconium-89 PET-CT for varied count levels
Where BSREM reconstruction is used for Zr-89 images, careful attention must be paid to the choice of weighting factor, especially for quantitative clinical studies. The effect of varying beta on several measures of image quality was characterised for the case of a phantom, with the results indicating that the value of beta for optimal Zr-89 quantitation is lower than what is generally used for optimal visualisation. This work demonstrates the need for careful attention to the reconstruction methods used for quantitative imaging studies, such as those required for theragnostic imaging.