Measuring Magnetization Curves of Ferrofluids: Comparison of the Differential Sweep Method and a Vibrating-Sample Magnetometer

A setup design is proposed for measuring magnetization curves by the differential sweep method. On the described setup, a number of magnetization curves were measured for both classical magnetic fluids and samples obtained by their solidification. The measurement results are compared with the results obtained using a vibrating-sample magnetometer. It is shown that the vibrating-sample magnetometer gives significantly lower values compared to the differential method when measuring the magnetization of a magnetic fluid. When measuring the magnetization of a cured magnetic-fluid sample, a complete coincidence of the results obtained by both measurement methods is observed. The discovered phenomenon is explained by the aggregation of particles in the magnetic fluid under the influence of a magnetic field. Estimations of the relative delay in the movement of aggregates are consistent with the observed divergence in the magnetization magnitude.

Authors

Institutions

Publication Details

Journal
Instruments and Experimental Techniques
Published
2024-10-01
DOI
https://doi.org/10.1134/s0020441224701574
Citations
2
Primary Topic
Characterization and Applications of Magnetic Nanoparticles
Type
article
Field-Weighted Citation Impact
0.37
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Measuring Magnetization Curves of Ferrofluids: Comparison of the Differential Sweep Method and a Vibrating-Sample Magnetometer

A. V. Lebedev
2 citations
Instruments and Experimental Techniques
Characterization and Applications of Magnetic Nanoparticles
0.37
article

Measuring Magnetization Curves of Ferrofluids: Comparison of the Differential Sweep Method and a Vibrating-Sample Magnetometer

A. V. Lebedev
article en
2 citations

Abstract

A setup design is proposed for measuring magnetization curves by the differential sweep method. On the described setup, a number of magnetization curves were measured for both classical magnetic fluids and samples obtained by their solidification. The measurement results are compared with the results obtained using a vibrating-sample magnetometer. It is shown that the vibrating-sample magnetometer gives significantly lower values compared to the differential method when measuring the magnetization of a magnetic fluid. When measuring the magnetization of a cured magnetic-fluid sample, a complete coincidence of the results obtained by both measurement methods is observed. The discovered phenomenon is explained by the aggregation of particles in the magnetic fluid under the influence of a magnetic field. Estimations of the relative delay in the movement of aggregates are consistent with the observed divergence in the magnetization magnitude.

Instruments and Experimental TechniquesVol. 67(5)
Institute of Continuous Media Mechanics (RU)
Openalex Percentile: Top 41%
Characterization and Applications of Magnetic Nanoparticles
0.37
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Measuring Magnetization Curves of Ferrofluids: Comparison of the Differential Sweep Method and a Vibrating-Sample Magnetometer — A. V. Lebedev · Instruments and Experimental Techniques (2024) | TGRS Research Map | TGRS