Bone Health Deterioration in Transfemoral Prosthetic Users: An Analytical Biomechanical Explanation

There is a five-decade recorded history indicating that persons with transfemoral amputation experience bone loss in their amputated femur at levels seen in bedridden and post-menopausal individuals, irrespective of age or mobility levels. We used computer simulation to recreate the mechanical environment created by the mechanical design of a prosthetic device in the surviving femur of individuals with transfemoral amputations. Finite element models of gait instances were developed from the hip joint computerized tomography scan of a subject along with a coupled ischial containment prosthetic socket fitted as per standard clinical guidelines. Accompanying mirror models, assembled similarly but without the prosthetic socket were used for stimulus comparison. Simulation showed that more than 90% of the trabecular bone volume in the amputated femur with an ischial containment socket registered compressive strain magnitudes below 300με. These strain magnitudes are below the threshold for bone maintenance as per mechanotransduction theory (i.e., they lie within the disuse window). Only 50% of the bone was in the disuse window for the mirror model for the gait instances considered. These results are consistent with reported in vivo evidence which shows that transfemoral prosthesis users may lose bone mass irrespective of age or mobility levels when using traditional socket designs. Clinically, this study shows that prosthetic sockets that support load through the ischium alter the kinetic chain and preclude application of mechanical stimulus that sustains healthy levels of bone mass in the proximal femur. The study also shows that femur length, prosthetic alignment and tissue tone influence this stimulus.

Authors

Institutions

Publication Details

Journal
International Journal for Numerical Methods in Biomedical Engineering
Published
2025-02-01
DOI
https://doi.org/10.1002/cnm.70014
Citations
2
Primary Topic
Prosthetics and Rehabilitation Robotics
Type
article
Field-Weighted Citation Impact
1.35

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Bone Health Deterioration in Transfemoral Prosthetic Users: An Analytical Biomechanical Explanation

Jose L. Zavaleta‐Ruiz, Matthew J. Major, Pankaj Pankaj
2 citations
International Journal for Numerical Methods in Biomedical Engineering
Prosthetics and Rehabilitation Robotics
1.35
article

Bone Health Deterioration in Transfemoral Prosthetic Users: An Analytical Biomechanical Explanation

Jose L. Zavaleta‐Ruiz, Matthew J. Major, Pankaj Pankaj
article en
2 citations

Abstract

There is a five-decade recorded history indicating that persons with transfemoral amputation experience bone loss in their amputated femur at levels seen in bedridden and post-menopausal individuals, irrespective of age or mobility levels. We used computer simulation to recreate the mechanical environment created by the mechanical design of a prosthetic device in the surviving femur of individuals with transfemoral amputations. Finite element models of gait instances were developed from the hip joint computerized tomography scan of a subject along with a coupled ischial containment prosthetic socket fitted as per standard clinical guidelines. Accompanying mirror models, assembled similarly but without the prosthetic socket were used for stimulus comparison. Simulation showed that more than 90% of the trabecular bone volume in the amputated femur with an ischial containment socket registered compressive strain magnitudes below 300με. These strain magnitudes are below the threshold for bone maintenance as per mechanotransduction theory (i.e., they lie within the disuse window). Only 50% of the bone was in the disuse window for the mirror model for the gait instances considered. These results are consistent with reported in vivo evidence which shows that transfemoral prosthesis users may lose bone mass irrespective of age or mobility levels when using traditional socket designs. Clinically, this study shows that prosthetic sockets that support load through the ischium alter the kinetic chain and preclude application of mechanical stimulus that sustains healthy levels of bone mass in the proximal femur. The study also shows that femur length, prosthetic alignment and tissue tone influence this stimulus.

International Journal for Numerical Methods in Biomedical EngineeringVol. 41(2)
Northwestern University (US), Jesse Brown VA Medical Center (US), University of Edinburgh (GB)
Consejo Nacional de Ciencia y Tecnología
Openalex Percentile: Top 22%
Prosthetics and Rehabilitation Robotics
1.35
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.