Long-term timing stabilization for pump–probe experiments at SACLA
We have developed a timing control system to stabilize the long-term timing drift between X-ray free-electron laser (XFEL) and optical laser pulses using an out-of-loop balanced optical-microwave phase detector and an arrival-timing monitor for pump-probe experiments at the SPring-8 Ångstrom Compact free-electron LAser (SACLA). The timing jitter and drift between the XFEL and the optical laser pulses have been reduced to less than 50 fs (RMS) over ∼49 h. The performance of the timing stabilization system was investigated by measuring the correlation of the long-term simultaneous timing monitoring on two branches of BL3 over 8 h. A linear correlation was observed with an RMS error of 8.6 fs.
Authors
- T. Yabuuchi (ORCID: https://orcid.org/0000-0003-0638-7640)
- Shigeki Owada (ORCID: https://orcid.org/0000-0002-1451-7612)
- Tadashi Togashi (ORCID: https://orcid.org/0000-0001-7615-7213)
- Makina Yabashi (ORCID: https://orcid.org/0000-0002-2472-1684)
Institutions
- SPring-8 (JP)
- Japan Synchrotron Radiation Research Institute (JP)
Publication Details
- Journal
- Journal of Synchrotron Radiation
- Published
- 2025-02-03
- DOI
- https://doi.org/10.1107/s1600577524011974
- Primary Topic
- Advanced X-ray Imaging Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00