Determination of carbon dioxide by gas chromatography using an electron capture detector for the analysis of greenhouse gases: A comparison and validation with the standard method

This study reports the development of a novel analytical method for the determination of carbon dioxide (CO<sub>2</sub>) along with the other main greenhouse gases (GHG)‒methane (CH<sub>4</sub>) and nitrous oxide (N<sub>2</sub>O)‒ from environmental samples by gas chromatography (GC) using an electron capture detector (ECD). This new method was compared and validated with the standard GC method for the determination of CO<sub>2</sub> using a thermal conductivity detector (TCD). The main performance parameters considered for the validation of an analytical method were evaluated: selectivity/specificity, linearity/working range, precision, trueness, limit of detection (LOD) and limit of quantitation (LOQ). This comparison was carried out using gas standards, reference materials and environmental samples containing GHG. Both the ECD- and TCD-based chromatographic analytical methods displayed a similar precision (3.1-3.4 %) and accuracy (101-106 %) for CO<sub>2</sub> analysis. Although the LOQ for the ECD detector is higher than that of the TCD detector (300 vs. 99 µmol mol⁻¹), it is still sufficient for the analysis of most GHG samples. Once validated, the new analytical method was used for the simultaneous determination of CO<sub>2</sub>, CH<sub>4</sub> and N<sub>2</sub>O in gaseous samples obtained from a wide range of conditions and agricultural environments. To this end, a GC instrument was equipped with a flame ionization detector (FID) for the determination of CH<sub>4</sub> and an ECD for CO<sub>2</sub> and N<sub>2</sub>O, connected in series by a system of valves. In this way, it was possible to accurately measure the main GHG present in the atmosphere and in gaseous samples, simplifying the required laboratory equipment and reducing the associated labor and costs.

Authors

Institutions

Publication Details

Journal
Journal of Chromatography A
Published
2025-02-03
DOI
https://doi.org/10.1016/j.chroma.2025.465750
Citations
8
Primary Topic
Analytical Chemistry and Chromatography
Type
article
Field-Weighted Citation Impact
6.05

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Determination of carbon dioxide by gas chromatography using an electron capture detector for the analysis of greenhouse gases: A comparison and validation with the standard method

August Bonmatí Blasi, Joan Noguerol Arias, Francesc X. Prenafeta‐Boldú, Míriam Cerrillo
8 citations
Journal of Chromatography A
Analytical Chemistry and Chromatography
6.05
article

Determination of carbon dioxide by gas chromatography using an electron capture detector for the analysis of greenhouse gases: A comparison and validation with the standard method

August Bonmatí Blasi, Joan Noguerol Arias, Francesc X. Prenafeta‐Boldú, Míriam Cerrillo
article en
8 citations

Abstract

This study reports the development of a novel analytical method for the determination of carbon dioxide (CO<sub>2</sub>) along with the other main greenhouse gases (GHG)‒methane (CH<sub>4</sub>) and nitrous oxide (N<sub>2</sub>O)‒ from environmental samples by gas chromatography (GC) using an electron capture detector (ECD). This new method was compared and validated with the standard GC method for the determination of CO<sub>2</sub> using a thermal conductivity detector (TCD). The main performance parameters considered for the validation of an analytical method were evaluated: selectivity/specificity, linearity/working range, precision, trueness, limit of detection (LOD) and limit of quantitation (LOQ). This comparison was carried out using gas standards, reference materials and environmental samples containing GHG. Both the ECD- and TCD-based chromatographic analytical methods displayed a similar precision (3.1-3.4 %) and accuracy (101-106 %) for CO<sub>2</sub> analysis. Although the LOQ for the ECD detector is higher than that of the TCD detector (300 vs. 99 µmol mol⁻¹), it is still sufficient for the analysis of most GHG samples. Once validated, the new analytical method was used for the simultaneous determination of CO<sub>2</sub>, CH<sub>4</sub> and N<sub>2</sub>O in gaseous samples obtained from a wide range of conditions and agricultural environments. To this end, a GC instrument was equipped with a flame ionization detector (FID) for the determination of CH<sub>4</sub> and an ECD for CO<sub>2</sub> and N<sub>2</sub>O, connected in series by a system of valves. In this way, it was possible to accurately measure the main GHG present in the atmosphere and in gaseous samples, simplifying the required laboratory equipment and reducing the associated labor and costs.

Journal of Chromatography AVol. 1745
Institute of Agrifood Research and Technology (ES)
Ministerio de Ciencia, Innovación y Universidades, Centres de Recerca de Catalunya, Generalitat de Catalunya
Openalex Percentile: Top 4%
Analytical Chemistry and Chromatography
6.05
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.