Evaluation of a modified formalism for precision electron beam dosimetry compared to TG-51 and TRS-398
This study supports the reliability of the modified formalism for electron beam reference dosimetry with chambers calibrated under <sup>60</sup>Co, thereby reducing uncertainty. The modified formalism yielded consistent results within smaller uncertainty (0.7 - 0.9 %). However, it is recommended to check each PPC's chamber-to-chamber variation.
Authors
- Y. K. Kim (ORCID: https://orcid.org/0000-0001-6426-7001)
- Rukundo Solomon (ORCID: https://orcid.org/0009-0000-5319-5807)
- Sang Hyun Choi (ORCID: https://orcid.org/0000-0002-6898-6617)
- Yonmo Sung (ORCID: https://orcid.org/0000-0001-6982-9974)
- Chul‐Young Yi (ORCID: https://orcid.org/0000-0003-4466-9710)
- In Jung Kim (ORCID: https://orcid.org/0000-0001-9161-6318)
- Jong In Park (ORCID: https://orcid.org/0000-0001-7059-3540)
- Seongmoon Jung (ORCID: https://orcid.org/0000-0002-9483-158X)
- Okky Agassy Firmansyah (ORCID: https://orcid.org/0000-0002-0068-5703)
Institutions
- Korea Research Institute of Standards and Science (KR)
- Korea Institute of Radiological and Medical Sciences (KR)
- Korea University of Science and Technology (KR)
Publication Details
- Journal
- Physica Medica
- Published
- 2025-02-01
- DOI
- https://doi.org/10.1016/j.ejmp.2025.104905
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Korea Research Institute of Standards and Science
- National Research Council of Science and Technology
- Ministry of Science and ICT, South Korea