Harmonic analysis and optimization for closed‐loop superconducting shim coils of 7 T MRI magnet
The study successfully demonstrates the creation of a set of second-order shim coils for a 7 T animal MRI system through numerical optimization. The design outputs provide essential technological support for the development of a human whole-body 7 T MRI system, ensuring high-quality imaging at the neuronal level. The project also highlights the importance of considering manufacturing and assembly flaws in the shim coil design process to achieve effective shimming in practical engineering scenarios.
Authors
- Jiaxin Li (ORCID: https://orcid.org/0009-0004-1492-1721)
- Lei Wu (ORCID: https://orcid.org/0000-0003-0572-5185)
- Panfeng Zhang (ORCID: https://orcid.org/0000-0001-6084-9231)
- Yongsuo Liu
- Chao Zhou (ORCID: https://orcid.org/0000-0002-6873-6140)
- Zhu-Ying Gao (ORCID: https://orcid.org/0009-0000-7342-8209)
- Zhan Zhang (ORCID: https://orcid.org/0000-0002-1754-8668)
- Zijie Lin (ORCID: https://orcid.org/0009-0000-1060-1916)
- Zhaohui Chu (ORCID: https://orcid.org/0000-0001-8320-7872)
Institutions
- University of Science and Technology of China (CN)
- Anhui University of Science and Technology (CN)
- Chinese Academy of Sciences (CN)
- Hefei Institutes of Physical Science (CN)
- Institute of Plasma Physics (CN)
Publication Details
- Journal
- Medical Physics
- Published
- 2025-01-28
- DOI
- https://doi.org/10.1002/mp.17641
- Citations
- 3
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 4.08
Funders
- Collaborative Innovation Project of Colleges and Universities of Anhui Province