Inappropriate Shocks in Brugada Syndrome Patients With a Subcutaneous Implantable Cardioverter Defibrillator
No significant difference was observed in the overall incidence of IAS between the S-ICD and TV-ICD groups; however, IAS due to non-AT causes occurred more frequently in the S-ICD group. Fewer ECG sensing vectors passing screening were significantly correlated with the IAS when exercise test results were considered. Therefore, ECG screening, including exercise testing, is preferable for S-ICD candidates.
Authors
- Tadafumi Nanbu (ORCID: https://orcid.org/0000-0002-7590-5029)
- Toshihisa Anzai (ORCID: https://orcid.org/0000-0003-3319-4967)
- Yoshitaka Tanaka (ORCID: https://orcid.org/0009-0001-9887-1503)
- George Suzuki (ORCID: https://orcid.org/0000-0001-8235-4221)
- Yuki Ishidoya (ORCID: https://orcid.org/0000-0003-2825-8479)
- Izumi Yoshida (ORCID: https://orcid.org/0000-0001-8921-052X)
- Tomomi Kanno
- Akihiko Yotsukura
- Masaya Watanabe
- Masayuki Sakurai
- Kazushige Inoue
- Junko Mitsuishi
Institutions
- Hokkaido University (JP)
- Hokko Memorial Hospital (JP)
Publication Details
- Journal
- Pacing and Clinical Electrophysiology
- Published
- 2025-01-27
- DOI
- https://doi.org/10.1111/pace.15151
- Citations
- 1
- Primary Topic
- Cardiac electrophysiology and arrhythmias
- Type
- article
- Field-Weighted Citation Impact
- 1.39