Novel anchor-type proximal femoral nail for the improvement of bone-fixation integrity in treating intertrochanteric fractures: an experimental and computational characterization study

This study introduces a novel anchor-type proximal femoral nail (AT-PFN) to improve the bone-fixation integrity over the standard screw-type nail (SST-PFN). Quasi-static incremental cyclic load test was performed to investigate load-displacement, cumulative deformation energy, time-strain, and backbone curves. The finite element analysis (FEA) was implemented to identify the stress and strain distributions. Additionally, non-destructive dynamic tests were conducted, and the measurements were processed using statistical pattern recognition, based on vector autoregression and principal component analysis to investigate the nonlinearity due to bone-fixation interface. The results demonstrate that the AT-PFN significantly improves the bone-fixation integrity compared to SST-PFN.

Authors

Institutions

Publication Details

Journal
Computer Methods in Biomechanics & Biomedical Engineering
Published
2025-01-27
DOI
https://doi.org/10.1080/10255842.2025.2456985
Citations
2
Primary Topic
Hip and Femur Fractures
Type
article
Field-Weighted Citation Impact
4.01
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Novel anchor-type proximal femoral nail for the improvement of bone-fixation integrity in treating intertrochanteric fractures: an experimental and computational characterization study

Mesut Tahta, Cemil Kayalı, Fırat Ozan, Mahmut Pekedis et al.
2 citations
Computer Methods in Biomechanics & Biomedical Engineering
Hip and Femur Fractures
4.01
article

Novel anchor-type proximal femoral nail for the improvement of bone-fixation integrity in treating intertrochanteric fractures: an experimental and computational characterization study

Mesut Tahta, Cemil Kayalı, Fırat Ozan, Mahmut Pekedis, Ahmet Adnan Karaarslan
article en
2 citations

Abstract

This study introduces a novel anchor-type proximal femoral nail (AT-PFN) to improve the bone-fixation integrity over the standard screw-type nail (SST-PFN). Quasi-static incremental cyclic load test was performed to investigate load-displacement, cumulative deformation energy, time-strain, and backbone curves. The finite element analysis (FEA) was implemented to identify the stress and strain distributions. Additionally, non-destructive dynamic tests were conducted, and the measurements were processed using statistical pattern recognition, based on vector autoregression and principal component analysis to investigate the nonlinearity due to bone-fixation interface. The results demonstrate that the AT-PFN significantly improves the bone-fixation integrity compared to SST-PFN.

Computer Methods in Biomechanics & Biomedical Engineering
Ege University (TR), Kayseri Eğitim ve Araştırma Hastanesi (TR), Izmir Bozyaka Eğitim ve Araştırma Hastanesi (TR)
Zero hunger
Openalex Percentile: Top 9%
Hip and Femur Fractures
4.01
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.