In Vivo Tissue Temperature Characteristics of Contact Force Catheter With a Mesh‐Shaped Irrigation Tip: A Porcine Study
HPSD ablation produced a modestly higher and more rapid increase in the tissue temperature around the veno-atrial junction with a shorter catheter tip-to-thermocouple distance required to reach lethal temperatures. This data contributes to understanding effective lesion creation and collateral injury prevention with the TactiFlex catheter.
Authors
- Rie Takahashi (ORCID: https://orcid.org/0000-0003-0255-7040)
- Masanaru Sawada (ORCID: https://orcid.org/0000-0002-9663-7617)
- Koichi Nagashima (ORCID: https://orcid.org/0000-0003-2326-225X)
- Hiroyuki Hao (ORCID: https://orcid.org/0000-0001-9951-0251)
- Yasuo Okumura (ORCID: https://orcid.org/0000-0002-2960-4241)
- Moyuru Hirata (ORCID: https://orcid.org/0000-0003-2701-486X)
- Yuji Wakamatsu (ORCID: https://orcid.org/0000-0002-1540-8514)
- Naoto Otsuka
- Kenta Uto (ORCID: https://orcid.org/0000-0003-2882-8785)
- Satoshi Hayashida (ORCID: https://orcid.org/0000-0002-8240-6201)
- Ryuta Watanabe (ORCID: https://orcid.org/0000-0003-2664-7139)
- Yoshiki Taniguchi (ORCID: https://orcid.org/0000-0002-7052-3187)
- Sayaka Kurokawa (ORCID: https://orcid.org/0000-0002-7989-3985)
- Shu Hirata
Institutions
- Nihon University (JP)
Publication Details
- Journal
- Pacing and Clinical Electrophysiology
- Published
- 2025-01-27
- DOI
- https://doi.org/10.1111/pace.15147
- Primary Topic
- Cardiac Arrhythmias and Treatments
- Type
- article
- Field-Weighted Citation Impact
- 0.00