Development of Simple Algorithm to Detect and Filter Motion Artifact Noise in Non Invasive Blood Pressure (NIBP) Measurement

Motion artifact is a noise induced by patient during NIBP measurement due to the patient’s intentional or unintentional movements. It can reduce the accuracy of the measurement. In this paper, we will conduct a study to identify, detect, and reduce the motion artifact effect to the NIBP measurement. MATLAB will be used to analyze the measurement signal, detecting and reducing the motion artifact present on the data. Our experiment shows that measurement data corrupted by motion artifacts has larger peaks in their frequency response. To reduce this noise, the noisy measurement signal is fit into a straight line, then the noisy measurement signal is subtracted with this straight line. The method has decreased MAP error by 67.48%, SYS error by 39.28%, and DIA error by 72.22%.

Authors

Institutions

Publication Details

Published
2024-12-03
DOI
https://doi.org/10.1109/apsipaasc63619.2025.10848895
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of Simple Algorithm to Detect and Filter Motion Artifact Noise in Non Invasive Blood Pressure (NIBP) Measurement

Trio Adiono, Clarence Amadeus, Sindy Novaria Cicilya Sinaga, Rd Elviana La’salina Muhlis
Non-Invasive Vital Sign Monitoring
article

Development of Simple Algorithm to Detect and Filter Motion Artifact Noise in Non Invasive Blood Pressure (NIBP) Measurement

Trio Adiono, Clarence Amadeus, Sindy Novaria Cicilya Sinaga, Rd Elviana La’salina Muhlis
article en

Abstract

Motion artifact is a noise induced by patient during NIBP measurement due to the patient’s intentional or unintentional movements. It can reduce the accuracy of the measurement. In this paper, we will conduct a study to identify, detect, and reduce the motion artifact effect to the NIBP measurement. MATLAB will be used to analyze the measurement signal, detecting and reducing the motion artifact present on the data. Our experiment shows that measurement data corrupted by motion artifacts has larger peaks in their frequency response. To reduce this noise, the noisy measurement signal is fit into a straight line, then the noisy measurement signal is subtracted with this straight line. The method has decreased MAP error by 67.48%, SYS error by 39.28%, and DIA error by 72.22%.

Bandung Institute of Technology (ID)
Openalex Percentile: Top 76%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.