Diagnostic Accuracy Comparison between Automatic and Conventional Blood Pressure Measuring Devices in Adults with Hypertension

Background Blood pressure measurement is a main clinical procedure used for the detection of Systemic Arterial Hypertension. Mercury sphygmomanometer, in combination with auscultation technique, is still recognized as the gold standard for non-invasive blood pressure measurement. Objective The objective of this study is to compare the diagnostic accuracy between automatic non-invasive blood pressure measurement devices versus mercury sphygmomanometers in hospitalized patients. Methods A prospective, analytical and cross-sectional diagnostic method study was designed to compare the concordance, positive and negative predictive value, the sensitivity and specificity of automatic non-invasive blood pressure measurement devices and the mercury sphygmomanometer. The devices were divided into two groups: the reference standard which was mercury sphygmomanometer (n=150), versus the index tests, that were wrist device (n=150), semi-automatic sphygmomanometer (n=150) and vital signs monitor (n=150). Results The monitor detected more cases (69%) of systemic arterial hypertension (p 0.0019, OR 5.3). The semiautomatic sphygmomanometer identified more true positives (n=22) and true negatives (n=113). It also showed a higher diagnostic accuracy: 67% sensitivity (p 0.0576, OR 0.5) and 97% specificity (p 0.088, OR 3.2). A mean arterial pressure variation <5mmHg of 65% was found with the semiautomatic sphygmomanometer ( p <0.0001, OR 0.02) and 61% with the vital signs monitor ( p <0.0001, OR 0.12). Best concordance of blood pressure between reference standard versus index test was obtained in the following order: Semi-automatic sphygmomanometer, Vital signs monitor, and Wrist device. Conclusion The use of vital signs monitor to measure blood pressure is recommended in all patients, although semiautomatic sphygmomanometer is more sensitive for the detection of systemic arterial hypertension.

Authors

Institutions

Publication Details

Journal
The Open Nursing Journal
Published
2025-01-27
DOI
https://doi.org/10.2174/0118744346375817250116104636
Primary Topic
Blood Pressure and Hypertension Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Diagnostic Accuracy Comparison between Automatic and Conventional Blood Pressure Measuring Devices in Adults with Hypertension

Omar Esteban Valencia-Ledezma, Nicolás Santiago-González, Rafael García-Rascón, Pedro Curi-Curi et al.
The Open Nursing Journal
Blood Pressure and Hypertension Studies
article

Diagnostic Accuracy Comparison between Automatic and Conventional Blood Pressure Measuring Devices in Adults with Hypertension

Omar Esteban Valencia-Ledezma, Nicolás Santiago-González, Rafael García-Rascón, Pedro Curi-Curi, Elia Hernandez-Roque, Evelin Y. Oliver-Flores, Guadalupe Curiel-Ramirez
article en

Abstract

Background Blood pressure measurement is a main clinical procedure used for the detection of Systemic Arterial Hypertension. Mercury sphygmomanometer, in combination with auscultation technique, is still recognized as the gold standard for non-invasive blood pressure measurement. Objective The objective of this study is to compare the diagnostic accuracy between automatic non-invasive blood pressure measurement devices versus mercury sphygmomanometers in hospitalized patients. Methods A prospective, analytical and cross-sectional diagnostic method study was designed to compare the concordance, positive and negative predictive value, the sensitivity and specificity of automatic non-invasive blood pressure measurement devices and the mercury sphygmomanometer. The devices were divided into two groups: the reference standard which was mercury sphygmomanometer (n=150), versus the index tests, that were wrist device (n=150), semi-automatic sphygmomanometer (n=150) and vital signs monitor (n=150). Results The monitor detected more cases (69%) of systemic arterial hypertension (p 0.0019, OR 5.3). The semiautomatic sphygmomanometer identified more true positives (n=22) and true negatives (n=113). It also showed a higher diagnostic accuracy: 67% sensitivity (p 0.0576, OR 0.5) and 97% specificity (p 0.088, OR 3.2). A mean arterial pressure variation <5mmHg of 65% was found with the semiautomatic sphygmomanometer ( p <0.0001, OR 0.02) and 61% with the vital signs monitor ( p <0.0001, OR 0.12). Best concordance of blood pressure between reference standard versus index test was obtained in the following order: Semi-automatic sphygmomanometer, Vital signs monitor, and Wrist device. Conclusion The use of vital signs monitor to measure blood pressure is recommended in all patients, although semiautomatic sphygmomanometer is more sensitive for the detection of systemic arterial hypertension.

The Open Nursing JournalVol. 19(1)
Mexican Social Security Institute (MX), Universidad México Contemporáneo (MX), Hospital Regional de Alta Especialidad del Bajío (MX)
Openalex Percentile: Top 98%
Blood Pressure and Hypertension Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.