Development of Two-Layer Depth-of-Interaction Detector for Small Animal Positron Emission Tomography(PET)/Magnetic Resonance Imaging (MRI) through Photoelectric Peak Analysis

A depth of interaction (DOI) detector was developed to improve the spatial resolution of positron emission tomography (PET) systems with magnetic resonance imaging (MRI) for small animals. The detector uses a silicon-based optical sensor that is not affected by magnetic fields and can be used in MRI systems. The design includes a two-layer scintillation pixel array with different reflectors for each layer such that the intensity and peak area of the signal detected by the optical sensor differs for each layer. A spectrum of the photoelectric peaks as a function of the energy was constructed by irradiating the scintillator block with Na-22 gamma radiation of 511 keV and by collecting the light generated by the silicon photomultiplier (SiPM) optical sensor. By analyzing the areas of the two photoelectric peaks separately, the DOI layer could be identified. The use of this detector is expected to enable excellent spatial resolution to be achieved in all areas of the field of view, and to produce images with excellent quality.

Authors

Publication Details

Journal
Journal of Magnetics
Published
2024-12-31
DOI
https://doi.org/10.4283/jmag.2024.29.4.575
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of Two-Layer Depth-of-Interaction Detector for Small Animal Positron Emission Tomography(PET)/Magnetic Resonance Imaging (MRI) through Photoelectric Peak Analysis

Seung‐Jae Lee
Journal of Magnetics
Medical Imaging Techniques and Applications
article

Development of Two-Layer Depth-of-Interaction Detector for Small Animal Positron Emission Tomography(PET)/Magnetic Resonance Imaging (MRI) through Photoelectric Peak Analysis

Seung‐Jae Lee
article en

Abstract

A depth of interaction (DOI) detector was developed to improve the spatial resolution of positron emission tomography (PET) systems with magnetic resonance imaging (MRI) for small animals. The detector uses a silicon-based optical sensor that is not affected by magnetic fields and can be used in MRI systems. The design includes a two-layer scintillation pixel array with different reflectors for each layer such that the intensity and peak area of the signal detected by the optical sensor differs for each layer. A spectrum of the photoelectric peaks as a function of the energy was constructed by irradiating the scintillator block with Na-22 gamma radiation of 511 keV and by collecting the light generated by the silicon photomultiplier (SiPM) optical sensor. By analyzing the areas of the two photoelectric peaks separately, the DOI layer could be identified. The use of this detector is expected to enable excellent spatial resolution to be achieved in all areas of the field of view, and to produce images with excellent quality.

Journal of MagneticsVol. 29(4)
Openalex Percentile: Top 57%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.