Optimizing Image Quality in Carbon Nanotube-Based Rectilinear Digital Tomosynthesis System via Projection Number Variation

External Beam Radiation Therapy (EBRT) is widely used in cancer treatment, requiring precise patient alignment for optimal outcomes. Traditional Cone-Beam Computed Tomography (CBCT) in Image-Guided Radiotherapy (IGRT) faces challenges like high radiation exposure and lengthy acquisition times. This study explores a rectilinear Digital Tomosynthesis System (DTS) with carbon nanotube (CNT)-based X-ray sources as an alternative for EBRT. By varying projection numbers, we assessed metrics including Structural Similarity Index Measure (SSIM), Signal-to-Noise Ratio (SNR), and Artifact Spread Function (ASF). Results indicated that increasing projection numbers enhanced SSIM, SNR, and artifact suppression, balancing image quality and computational efficiency. Integrating GPU acceleration further optimized processing speed, supporting real-time patient positioning in radiotherapy without conventional on-board imaging.

Authors

Publication Details

Journal
Journal of Magnetics
Published
2024-12-31
DOI
https://doi.org/10.4283/jmag.2024.29.4.480
Citations
1
Primary Topic
Advanced Electron Microscopy Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.52
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimizing Image Quality in Carbon Nanotube-Based Rectilinear Digital Tomosynthesis System via Projection Number Variation

Byungdu Jo
1 citations
Journal of Magnetics
Advanced Electron Microscopy Techniques and Applications
0.52
article

Optimizing Image Quality in Carbon Nanotube-Based Rectilinear Digital Tomosynthesis System via Projection Number Variation

Byungdu Jo
article en
1 citations

Abstract

External Beam Radiation Therapy (EBRT) is widely used in cancer treatment, requiring precise patient alignment for optimal outcomes. Traditional Cone-Beam Computed Tomography (CBCT) in Image-Guided Radiotherapy (IGRT) faces challenges like high radiation exposure and lengthy acquisition times. This study explores a rectilinear Digital Tomosynthesis System (DTS) with carbon nanotube (CNT)-based X-ray sources as an alternative for EBRT. By varying projection numbers, we assessed metrics including Structural Similarity Index Measure (SSIM), Signal-to-Noise Ratio (SNR), and Artifact Spread Function (ASF). Results indicated that increasing projection numbers enhanced SSIM, SNR, and artifact suppression, balancing image quality and computational efficiency. Integrating GPU acceleration further optimized processing speed, supporting real-time patient positioning in radiotherapy without conventional on-board imaging.

Journal of MagneticsVol. 29(4)
Openalex Percentile: Top 17%
Advanced Electron Microscopy Techniques and Applications
0.52
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Optimizing Image Quality in Carbon Nanotube-Based Rectilinear Digital Tomosynthesis System via Projection Number Variation — Byungdu Jo · Journal of Magnetics (2024) | TGRS Research Map | TGRS