Impact of a novel limbal-rigid contact lens on potential cost savings in Stevens-Johnson syndrome patients from postmarket surveillance

To investigate the safety, efficacy, and potential-cost-savings of a newly-developed limbal-rigid contact lens (CL) via postmarket surveillance (PMS). Postmarket prospective observational use-results surveillance with cost-savings analysis. We examined 80 eyes of 68 Stevens-Johnson syndrome (SJS) cases using the newly-developed Suncon Kyoto-CS limbal-rigid CL (Sun Contact Lens). CL-wear-associated best-corrected visual acuity (BCVA), objective signs (i.e., bulbar-conjunctival hyperemia, upper-eyelid conjunctival scarring, eye discharge, keratoconjunctival epithelial defect, corneal opacification, and corneal neovascularization), and subjective symptoms (i.e., dryness and eye pain) at baseline and at 6-months of CL use were measured and compared, with adverse events noted. Potential cost savings from improvements in CL-wear-related BCVA were estimated based on the potential for improvements in each patient's visual-impairment-grade under the Japanese Pharmaceuticals and Medical Devices Agency (PMDA) disability pension plan. Main outcome measures were potential cost savings via use of the CL. Secondary outcome measures were CL safety and efficacy via PMS. At 6-months of CL use, the median BCVA in the CL-fitted eyes improved from 1.30 to 0.70 logMAR compared to that of spectacle correction at baseline ( P < 0.001), and improvement of bulbar-conjunctival hyperemia, dryness, and eye pain was observed ( P < 0.001, P < 0.001, and P < 0.001, respectively). No serious adverse events occurred. Potential-cost-savings analysis showed visual-impairment-grade improvement in 18 (28.6 %) of 63 patients. Considering the financial amount of the PMDA disability pension, estimated total potential cost savings was >6.9-million U.S. dollars. The newly-developed limbal-rigid CL was found safe and effective for treating SJS-associated ocular sequelae, and potentially reduces disability-pension costs.

Authors

Institutions

Publication Details

Journal
AJO International
Published
2025-01-23
DOI
https://doi.org/10.1016/j.ajoint.2025.100101
Primary Topic
Drug-Induced Adverse Reactions
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of a novel limbal-rigid contact lens on potential cost savings in Stevens-Johnson syndrome patients from postmarket surveillance

Naoki Yamauchi, Chie Sotozono, Takefumi Yamaguchi, Satoshi Teramukai et al.
AJO International
Drug-Induced Adverse Reactions
article

Impact of a novel limbal-rigid contact lens on potential cost savings in Stevens-Johnson syndrome patients from postmarket surveillance

Naoki Yamauchi, Chie Sotozono, Takefumi Yamaguchi, Satoshi Teramukai, Hideki Fukuoka, Mayumi Ueta, Shigeru Kinoshita, Yoshiyuki Satake, Jun‐ichi Miyazaki, Tsutomu Inatomi, Jun Shimazaki, Satomi Sakabayashi, Kenji Konomi, Kenichi Kimura, Yoko Ogawa, Koji Koji, Kojiro Imai, Yuichi Uchino, Ryoichi Shiota
article en

Abstract

To investigate the safety, efficacy, and potential-cost-savings of a newly-developed limbal-rigid contact lens (CL) via postmarket surveillance (PMS). Postmarket prospective observational use-results surveillance with cost-savings analysis. We examined 80 eyes of 68 Stevens-Johnson syndrome (SJS) cases using the newly-developed Suncon Kyoto-CS limbal-rigid CL (Sun Contact Lens). CL-wear-associated best-corrected visual acuity (BCVA), objective signs (i.e., bulbar-conjunctival hyperemia, upper-eyelid conjunctival scarring, eye discharge, keratoconjunctival epithelial defect, corneal opacification, and corneal neovascularization), and subjective symptoms (i.e., dryness and eye pain) at baseline and at 6-months of CL use were measured and compared, with adverse events noted. Potential cost savings from improvements in CL-wear-related BCVA were estimated based on the potential for improvements in each patient's visual-impairment-grade under the Japanese Pharmaceuticals and Medical Devices Agency (PMDA) disability pension plan. Main outcome measures were potential cost savings via use of the CL. Secondary outcome measures were CL safety and efficacy via PMS. At 6-months of CL use, the median BCVA in the CL-fitted eyes improved from 1.30 to 0.70 logMAR compared to that of spectacle correction at baseline ( P < 0.001), and improvement of bulbar-conjunctival hyperemia, dryness, and eye pain was observed ( P < 0.001, P < 0.001, and P < 0.001, respectively). No serious adverse events occurred. Potential-cost-savings analysis showed visual-impairment-grade improvement in 18 (28.6 %) of 63 patients. Considering the financial amount of the PMDA disability pension, estimated total potential cost savings was >6.9-million U.S. dollars. The newly-developed limbal-rigid CL was found safe and effective for treating SJS-associated ocular sequelae, and potentially reduces disability-pension costs.

AJO InternationalVol. 2(1)
Keio University (JP), Kyoto Prefectural University of Medicine (JP), Tokyo Dental College Ichikawa General Hospital (JP), Keio University Hospital (JP), University Hospital Kyoto Prefectural University of Medicine (JP), Lens Technology (China) (CN)
Decent work and economic growth
Openalex Percentile: Top 99%
Drug-Induced Adverse Reactions
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.