Verification of tangential breast treatment dose during 2D treatment planning

This study presents a water phantom study of verification of tangential breast treatment dose calculations in a two-dimensional treatment planning system (TPS) at the Department of Radiotherapy, in a National hospital in Sri Lanka. Measurements were made using a 0.6 cc Farmer-type cylindrical ionization chamber with an electrometer in a breast phantom made of water field Perspex structure. The measured doses in the breast were compared with calculations made on a two-dimensional radiotherapy treatment planning software THERAPLAN®PLUS version 3.55 using the digitized breast phantom contour. Overall, the measurements obtained from the TPS had an accuracy of ±5% compared to the direct measurements while most of the measurements lie within ±3%. And dose in the chest wall was below the tolerance level of the lung and heart. Thus, it was concluded that TPS calculations and dose distributions are within the clinically acceptable accuracy level of international standards (ICRU report 50).

Authors

Publication Details

Journal
Sri Lankan Journal of Physics
Published
2024-12-31
DOI
https://doi.org/10.4038/sljp.v25i2.8135
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Verification of tangential breast treatment dose during 2D treatment planning

V.P.S. Perera, C. Jayalath, Velautham Sivakumar, Y.P.Y.P. Ariyasinghe et al.
Sri Lankan Journal of Physics
Advanced Radiotherapy Techniques
article

Verification of tangential breast treatment dose during 2D treatment planning

V.P.S. Perera, C. Jayalath, Velautham Sivakumar, Y.P.Y.P. Ariyasinghe, A. J. Hilmi
article en

Abstract

This study presents a water phantom study of verification of tangential breast treatment dose calculations in a two-dimensional treatment planning system (TPS) at the Department of Radiotherapy, in a National hospital in Sri Lanka. Measurements were made using a 0.6 cc Farmer-type cylindrical ionization chamber with an electrometer in a breast phantom made of water field Perspex structure. The measured doses in the breast were compared with calculations made on a two-dimensional radiotherapy treatment planning software THERAPLAN®PLUS version 3.55 using the digitized breast phantom contour. Overall, the measurements obtained from the TPS had an accuracy of ±5% compared to the direct measurements while most of the measurements lie within ±3%. And dose in the chest wall was below the tolerance level of the lung and heart. Thus, it was concluded that TPS calculations and dose distributions are within the clinically acceptable accuracy level of international standards (ICRU report 50).

Sri Lankan Journal of PhysicsVol. 25(2)
National Science Foundation
Openalex Percentile: Top 65%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Verification of tangential breast treatment dose during 2D treatment planning — V.P.S. Perera, C. Jayalath, et al. · Sri Lankan Journal of Physics (2024) | TGRS Research Map | TGRS