Systematical Theoretical Study of the Nonpoint Source Effects in Nuclear Medicine Shielding Calculation

The objective of this note was to systematically study the effects from nonpoint radiation sources on the nuclear medicine imaging room or uptake room shielding design. The dose from a nonpoint source to a location at a certain distance to the center of the source was calculated. A correction factor could be applied to the dose calculated using a point source model with the same distance to compensate the nonpoint source effects. The correction factor was a function only depending on the source relative dimension, the ratio of the source's geometric dimension to the distance. Theoretical and numerical calculation were performed to calculate the correction factor on a one-dimensional line source, two-dimensional circular disc source, three-dimensional sphere, and cylinder sources. Our study indicated that for large nuclear medicine imaging rooms, the point source model was a good approximation for shielding design. For a small uptake room, a nonpoint source correction may be considered to calculate the barrier thickness.

Authors

Institutions

Publication Details

Journal
Health Physics
Published
2025-01-16
DOI
https://doi.org/10.1097/hp.0000000000001930
Primary Topic
Nuclear Physics and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Systematical Theoretical Study of the Nonpoint Source Effects in Nuclear Medicine Shielding Calculation

Tianliang Gu
Health Physics
Nuclear Physics and Applications
article

Systematical Theoretical Study of the Nonpoint Source Effects in Nuclear Medicine Shielding Calculation

Tianliang Gu
article en

Abstract

The objective of this note was to systematically study the effects from nonpoint radiation sources on the nuclear medicine imaging room or uptake room shielding design. The dose from a nonpoint source to a location at a certain distance to the center of the source was calculated. A correction factor could be applied to the dose calculated using a point source model with the same distance to compensate the nonpoint source effects. The correction factor was a function only depending on the source relative dimension, the ratio of the source's geometric dimension to the distance. Theoretical and numerical calculation were performed to calculate the correction factor on a one-dimensional line source, two-dimensional circular disc source, three-dimensional sphere, and cylinder sources. Our study indicated that for large nuclear medicine imaging rooms, the point source model was a good approximation for shielding design. For a small uptake room, a nonpoint source correction may be considered to calculate the barrier thickness.

Health PhysicsVol. 128(5)
The University of Texas Health Science Center (US)
Openalex Percentile: Top 99%
Nuclear Physics and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Systematical Theoretical Study of the Nonpoint Source Effects in Nuclear Medicine Shielding Calculation — Tianliang Gu · Health Physics (2025) | TGRS Research Map | TGRS