Real-World Life Analysis of a Continuous Glucose Monitoring and Smart Insulin Pen System in Type 1 Diabetes: A Cohort Study

Background: Diabetes affects over 460 million people worldwide and represents a growing public health challenge driven largely by dietary and lifestyle factors. While Type 2 diabetes (T2D) is more prevalent, Type 1 diabetes (T1D) presents unique therapeutic challenges, particularly in younger individuals. Advances in diabetes management, such as continuous glucose monitoring (CGM), insulin pumps (IP), and, more recently, smart multiple dose injection (MDI) pens, have significantly enhanced glycemic control and improved patients’ quality of life. Aim: This study aims to evaluate the baseline characteristics of patients switching from MDI therapy to the Medtronic Smart MDI system [composed of a smart insulin pen (InPenTM) and a connected CGM Medtronic SimpleraTM sensor] and to assess its impact on glycemic outcomes over different time periods (14, 30, and 90 days). Methods: A retrospective observational study was conducted among adults with T1D who initiated Medtronic Smart MDI therapy. Participants were enrolled voluntarily at the Diabetes and Nutrition Clinic in Ast Fermo, Marche Region, Italy. Glycemic parameters were monitored using CGM data and analyzed with descriptive statistics, including mean, standard deviation (SD), and interquartile range (IQR). Comparisons across time periods were performed using the Wilcoxon signed-rank test, with statistical significance set at p < 0.05. Results: This study included 21 participants with a mean age of 51.5 years, a mean BMI of 24.7, and a mean duration of T1D of 21.9 years. The transition from a traditional MDI system to the Smart MDI system resulted in significant improvements in key glycemic parameters: mean Sensor Glucose (SG) decreased from 171.0 mg/dL to 153.5 mg/dL (p = 0.035), Time In Range (TIR) increased from 58.0% to 64.4% (p = 0.005), and time above range (TAR; >180 mg/dL) decreased from 39.0% to 34.2% (p = 0.015). No significant differences were observed in the time below range (TBR). Conclusions: The transition to the Medtronic Smart MDI system significantly enhanced glycemic control by lowering mean glucose levels and increasing TIR. These findings highlight its efficacy in improving hyperglycemia management while maintaining a stable risk of hypoglycemia.

Authors

Institutions

Publication Details

Journal
Diabetology
Published
2025-01-16
DOI
https://doi.org/10.3390/diabetology6010007
Citations
14
Primary Topic
Diabetes Management and Research
Type
article
Field-Weighted Citation Impact
18.13
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Real-World Life Analysis of a Continuous Glucose Monitoring and Smart Insulin Pen System in Type 1 Diabetes: A Cohort Study

Gabriele Caggianelli, Mauro Parozzi, Gaetano Ferrara, Stefano Mancin et al.
14 citations
Diabetology
Diabetes Management and Research
18.13
article

Real-World Life Analysis of a Continuous Glucose Monitoring and Smart Insulin Pen System in Type 1 Diabetes: A Cohort Study

Gabriele Caggianelli, Mauro Parozzi, Gaetano Ferrara, Stefano Mancin, Sara Morales Palomares, Marco Sguanci, Paola Pantanetti, Giovanni Cangelosi, Federico Biondini, Fabio Petrelli
article en
14 citations

Abstract

Background: Diabetes affects over 460 million people worldwide and represents a growing public health challenge driven largely by dietary and lifestyle factors. While Type 2 diabetes (T2D) is more prevalent, Type 1 diabetes (T1D) presents unique therapeutic challenges, particularly in younger individuals. Advances in diabetes management, such as continuous glucose monitoring (CGM), insulin pumps (IP), and, more recently, smart multiple dose injection (MDI) pens, have significantly enhanced glycemic control and improved patients’ quality of life. Aim: This study aims to evaluate the baseline characteristics of patients switching from MDI therapy to the Medtronic Smart MDI system [composed of a smart insulin pen (InPenTM) and a connected CGM Medtronic SimpleraTM sensor] and to assess its impact on glycemic outcomes over different time periods (14, 30, and 90 days). Methods: A retrospective observational study was conducted among adults with T1D who initiated Medtronic Smart MDI therapy. Participants were enrolled voluntarily at the Diabetes and Nutrition Clinic in Ast Fermo, Marche Region, Italy. Glycemic parameters were monitored using CGM data and analyzed with descriptive statistics, including mean, standard deviation (SD), and interquartile range (IQR). Comparisons across time periods were performed using the Wilcoxon signed-rank test, with statistical significance set at p < 0.05. Results: This study included 21 participants with a mean age of 51.5 years, a mean BMI of 24.7, and a mean duration of T1D of 21.9 years. The transition from a traditional MDI system to the Smart MDI system resulted in significant improvements in key glycemic parameters: mean Sensor Glucose (SG) decreased from 171.0 mg/dL to 153.5 mg/dL (p = 0.035), Time In Range (TIR) increased from 58.0% to 64.4% (p = 0.005), and time above range (TAR; >180 mg/dL) decreased from 39.0% to 34.2% (p = 0.015). No significant differences were observed in the time below range (TBR). Conclusions: The transition to the Medtronic Smart MDI system significantly enhanced glycemic control by lowering mean glucose levels and increasing TIR. These findings highlight its efficacy in improving hyperglycemia management while maintaining a stable risk of hypoglycemia.

DiabetologyVol. 6(1)
Azienda Ospedaliera San Giovanni Addolorata (IT), Istituto Ramazzini (IT), Ospedale Policlinico San Martino (IT), IRCCS Humanitas Research Hospital (IT), University of Calabria (IT)
Openalex Percentile: Top 1%
Diabetes Management and Research
18.13
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.