Accuracy of flow volume estimation in the dilated aorta using 4D flow MRI: a pulsatile phantom study

<i>Objectives.</i>Aortic dilatation is a severe pathology that increases the risk of rupture and its hemodynamics could be accurately assessed by using the 4D flow cardiovascular magnetic resonance (CMR) technique but flow assessment under complex flow patterns require validation. The aim of this work was to develop an<i>in vitro</i>system compatible with CMR to assess the accuracy of volume flow measurements in dilated aortas.<i>Approach.</i>Two latex models, one with ascending and the other with abdominal aortic aneurysms were manufactured to ensure a constant and controlled net flow volume along the aortic length. A pneumatic piston driven by a stepper motor and controlled by an embedded system located in the control room modulated a pulsatile fluid flow using a pump with an elastic membrane placed in the magnet near the elastic models. All the visualization and measurement algorithms were integrated into a custom computer platform. 4D flow imaging was used to estimate the flow rate and volume through multiple aortic planes and compared to the reference assessed by weight method and to 2D flow measurements.<i>Main results.</i>The errors of flow volume assessment using 4D flow remained within reasonable limits along the length of the aortic models. Mean differences in net flow volume from the reference were less than 2 ml (range -4 to 6 ml), corresponding to mean relative differences of less than 4% (range -8% to 11%). Averaged net, forward and backward flow volume estimations along the aortic length were similar using 2D and 4D flow measurements (<i>p</i>> 0.05). Peak forward and backward flow rates increased in the dilated regions and were comparable to those observed in patients.<i>Significance.</i>The accuracy of flow volume estimates in complex flow patterns, such as those observed in patients with aneurysms, was validated<i>in vitro</i>using 4D flow.

Authors

Institutions

Publication Details

Journal
Physiological Measurement
Published
2025-01-16
DOI
https://doi.org/10.1088/1361-6579/adab4e
Citations
3
Primary Topic
Aortic aneurysm repair treatments
Type
article
Field-Weighted Citation Impact
4.56

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy of flow volume estimation in the dilated aorta using 4D flow MRI: a pulsatile phantom study

A.A. Valda, Damián Craiem, Mariano E. Casciaro, Eduardo E. Rodríguez et al.
3 citations
Physiological Measurement
Aortic aneurysm repair treatments
4.56
article

Accuracy of flow volume estimation in the dilated aorta using 4D flow MRI: a pulsatile phantom study

A.A. Valda, Damián Craiem, Mariano E. Casciaro, Eduardo E. Rodríguez, Sebastián Graf, Edmundo I. Cabrera Fischer
article en
3 citations

Abstract

<i>Objectives.</i>Aortic dilatation is a severe pathology that increases the risk of rupture and its hemodynamics could be accurately assessed by using the 4D flow cardiovascular magnetic resonance (CMR) technique but flow assessment under complex flow patterns require validation. The aim of this work was to develop an<i>in vitro</i>system compatible with CMR to assess the accuracy of volume flow measurements in dilated aortas.<i>Approach.</i>Two latex models, one with ascending and the other with abdominal aortic aneurysms were manufactured to ensure a constant and controlled net flow volume along the aortic length. A pneumatic piston driven by a stepper motor and controlled by an embedded system located in the control room modulated a pulsatile fluid flow using a pump with an elastic membrane placed in the magnet near the elastic models. All the visualization and measurement algorithms were integrated into a custom computer platform. 4D flow imaging was used to estimate the flow rate and volume through multiple aortic planes and compared to the reference assessed by weight method and to 2D flow measurements.<i>Main results.</i>The errors of flow volume assessment using 4D flow remained within reasonable limits along the length of the aortic models. Mean differences in net flow volume from the reference were less than 2 ml (range -4 to 6 ml), corresponding to mean relative differences of less than 4% (range -8% to 11%). Averaged net, forward and backward flow volume estimations along the aortic length were similar using 2D and 4D flow measurements (<i>p</i>> 0.05). Peak forward and backward flow rates increased in the dilated regions and were comparable to those observed in patients.<i>Significance.</i>The accuracy of flow volume estimates in complex flow patterns, such as those observed in patients with aneurysms, was validated<i>in vitro</i>using 4D flow.

Physiological MeasurementVol. 46(1)
Consejo Nacional de Investigaciones Científicas y Técnicas (AR), Favaloro University (AR), National University of General San Martín (AR), National University of General Sarmiento (AR)
Agencia Nacional de Promoción Científica y Tecnológica
Good health and well-being
Openalex Percentile: Top 7%
Aortic aneurysm repair treatments
4.56
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.