Total fluorine analysis in solid samples by instrumental neutron activation analysis coupled with the internal standard method

• Mass fraction of F quantified via instrumental neutron activation analysis coupled with internal standard. • Normalized relation between F amount and 20 F counting rate does not vary over time. • Mass fraction of F can be quantified using a robust and linear calibration curve. • Analytical performance validated using certified reference material of high-putiy organic compound and SiC. • Proposed method used to quantify the mass fraction of F in synthetic polymer resin. Instrumental neutron activation analysis (INAA) is a valuable technique for quantifying F; however, it is associated with significant measurement uncertainty, which can be reduced by combining INAA with the internal standard method. In this study, known amounts of Sc were added to both samples and calibration standards to compensate for variations in the fluence rates of the irradiated neutrons and measurements of the γ-rays. The 46m Sc produced during neutron irradiation was used as the internal standard. Data were collected using two approaches: a single irradiation for 5 s followed by a 10 s γ-ray measurement, and a cyclic method involving five repetitions of the single irradiation and γ-ray measurements. Normalization of the counting rate using the internal standard effectively eliminated day-to-day variations, and a reliable relationship was established between the normalized signal and the amount of F. The proposed method produced a robust and highly linear calibration plot, enabling the accurate quantification of the mass fraction of F. The analytical performance of the proposed method was validated using certified reference materials before using it to quantify the mass fraction of F in commercially available synthetic polymer resins. The cyclic irradiation method, which required approximately 30 min for measurement, yielded a lower expanded uncertainty in the mass fraction of F compared with the single irradiation method, in which measurements were completed within 30 s. Consequently, the single irradiation method was more suitable for screening, whereas the cyclic irradiation method was more suitable for the precise quantification of F.

Authors

Institutions

Publication Details

Journal
Talanta Open
Published
2025-01-16
DOI
https://doi.org/10.1016/j.talo.2025.100409
Citations
3
Primary Topic
Nuclear Physics and Applications
Type
article
Field-Weighted Citation Impact
5.65
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Total fluorine analysis in solid samples by instrumental neutron activation analysis coupled with the internal standard method

Tsutomu Miura
3 citations
Talanta Open
Nuclear Physics and Applications
5.65
article

Total fluorine analysis in solid samples by instrumental neutron activation analysis coupled with the internal standard method

Tsutomu Miura
article en
3 citations

Abstract

• Mass fraction of F quantified via instrumental neutron activation analysis coupled with internal standard. • Normalized relation between F amount and 20 F counting rate does not vary over time. • Mass fraction of F can be quantified using a robust and linear calibration curve. • Analytical performance validated using certified reference material of high-putiy organic compound and SiC. • Proposed method used to quantify the mass fraction of F in synthetic polymer resin. Instrumental neutron activation analysis (INAA) is a valuable technique for quantifying F; however, it is associated with significant measurement uncertainty, which can be reduced by combining INAA with the internal standard method. In this study, known amounts of Sc were added to both samples and calibration standards to compensate for variations in the fluence rates of the irradiated neutrons and measurements of the γ-rays. The 46m Sc produced during neutron irradiation was used as the internal standard. Data were collected using two approaches: a single irradiation for 5 s followed by a 10 s γ-ray measurement, and a cyclic method involving five repetitions of the single irradiation and γ-ray measurements. Normalization of the counting rate using the internal standard effectively eliminated day-to-day variations, and a reliable relationship was established between the normalized signal and the amount of F. The proposed method produced a robust and highly linear calibration plot, enabling the accurate quantification of the mass fraction of F. The analytical performance of the proposed method was validated using certified reference materials before using it to quantify the mass fraction of F in commercially available synthetic polymer resins. The cyclic irradiation method, which required approximately 30 min for measurement, yielded a lower expanded uncertainty in the mass fraction of F compared with the single irradiation method, in which measurements were completed within 30 s. Consequently, the single irradiation method was more suitable for screening, whereas the cyclic irradiation method was more suitable for the precise quantification of F.

Talanta OpenVol. 11
National Institute of Advanced Industrial Science and Technology (JP)
Openalex Percentile: Top 5%
Nuclear Physics and Applications
5.65
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.