Challenging Extraction and Replacement of an Eight-year-old Nanostim Leadless Pacemaker and AVEIR Implant

Leadless pacemakers (LPs) are emerging options for bradyarrhythmias. However, extraction can be risky if the device is in an unfavorable position. We present a challenging case of a Nanostim LP (NLP) (Abbott Medical Inc., Abbott Park, IL, USA) placed 8 years prior to removal and subsequent replacement with an AVEIR LP (ALP) (Abbott). A 72-year-old man received an NLP in 2015 for persistent atrial fibrillation with bradycardia. The NLP could not be interrogated in our office. An external event monitor demonstrated persistent atrial fibrillation with bradycardia and pauses. A premature battery depletion of the NLP was suspected. An ALP was chosen for replacement. On a computed tomography scan of the chest, the NLP was seen in the mid-free wall of the right ventricle, and its proximal portion abutted the tricuspid annulus. The AVEIR retrieval catheter (ARC) was used for retrieval. Multiple attempts were made to snare the device, yet it proved difficult due to poor placement and button tissue formation. The snare became damaged, and a second ARC was needed. On the successful attempt, the NLP was snared proximally and permitted docking. We advanced the protective sleeve over the NLP body, but resistance was observed due to tissue growth. Counterclockwise torsion was applied, and the device disconnected. Once the NLP was in linear orientation, the protective sleeve was eventually positioned over its body, allowing removal. The ALP was then installed without difficulty or structural complications. This report shows the importance of proper LP placement in the right ventricular septal wall. LP removal can be performed safely, yet complications can arise based on the age and location of the LP. The ARC can be successfully used to extract non-AVEIR LPs with evidence of docking button tissue growth. Similar interventions should exercise caution when attempting extraction and subsequent implantation.

Authors

Institutions

Publication Details

Journal
Journal of Innovations in Cardiac Rhythm Management
Published
2024-12-01
DOI
https://doi.org/10.19102/icrm.2024.15126
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Challenging Extraction and Replacement of an Eight-year-old Nanostim Leadless Pacemaker and AVEIR Implant

Rajiv A. Kabadi, Binaya Basyal, Evan Czulada, Cyrus A. Hadadi et al.
Journal of Innovations in Cardiac Rhythm Management
Cardiac pacing and defibrillation studies
article

Challenging Extraction and Replacement of an Eight-year-old Nanostim Leadless Pacemaker and AVEIR Implant

Rajiv A. Kabadi, Binaya Basyal, Evan Czulada, Cyrus A. Hadadi, Athanasios Thomaides
article en

Abstract

Leadless pacemakers (LPs) are emerging options for bradyarrhythmias. However, extraction can be risky if the device is in an unfavorable position. We present a challenging case of a Nanostim LP (NLP) (Abbott Medical Inc., Abbott Park, IL, USA) placed 8 years prior to removal and subsequent replacement with an AVEIR LP (ALP) (Abbott). A 72-year-old man received an NLP in 2015 for persistent atrial fibrillation with bradycardia. The NLP could not be interrogated in our office. An external event monitor demonstrated persistent atrial fibrillation with bradycardia and pauses. A premature battery depletion of the NLP was suspected. An ALP was chosen for replacement. On a computed tomography scan of the chest, the NLP was seen in the mid-free wall of the right ventricle, and its proximal portion abutted the tricuspid annulus. The AVEIR retrieval catheter (ARC) was used for retrieval. Multiple attempts were made to snare the device, yet it proved difficult due to poor placement and button tissue formation. The snare became damaged, and a second ARC was needed. On the successful attempt, the NLP was snared proximally and permitted docking. We advanced the protective sleeve over the NLP body, but resistance was observed due to tissue growth. Counterclockwise torsion was applied, and the device disconnected. Once the NLP was in linear orientation, the protective sleeve was eventually positioned over its body, allowing removal. The ALP was then installed without difficulty or structural complications. This report shows the importance of proper LP placement in the right ventricular septal wall. LP removal can be performed safely, yet complications can arise based on the age and location of the LP. The ARC can be successfully used to extract non-AVEIR LPs with evidence of docking button tissue growth. Similar interventions should exercise caution when attempting extraction and subsequent implantation.

Journal of Innovations in Cardiac Rhythm ManagementVol. 15(12)
Georgetown University (US)
Openalex Percentile: Top 55%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.