Introduction: The primary stability and insertion torque of osseintegrated implants is fundamental to the success of osseointegration, especially in cases where a prosthesis with immediate load is installed. Objetive: This study aimed to assess the influence of the macrogeometry of two implant types on insertion torque and primary stability. Materials and Methods: This study utilized 40 implants, comprising 20 conical and 20 cylindrical, divided into four experimental groups. These implants were installed by a single operator, in a synthetic bone block following the manufacturer's recommendations and the sub-instrumentation technique. Implant stability was immediately assessed post-installation using magnetic resonance frequency analysis, measured by Osstellâ. The insertion torque was determined using a surgical implant installation motor. The resulting data were subjected to statistical analysis using the Kruskal–Wallis test, with Dunn's post hoc test employed to determine any significant differences between the implant groups in terms of insertion torque and primary stability. Results: A significant difference was observed in the stability of buccolingual (p=0.002) and the average stability (p=0.008), with respective effect sizes of 0.197 and 0.335. The insertion torque also demonstrated significance (p=0.018). Notably, a decrease in the values of the 04×10 implants was observed when compared to the 3.5×10 implants. Conclusions: Implants with a smaller diameter exhibited greater primary stability, whereas those with a larger diameter demonstrated higher insertion torque values. Keywords: Dental implant; primary stability; insertion torque.

Authors

Institutions

Publication Details

Journal
Revista de Ciências Médicas e Biológicas
Published
2025-01-09
DOI
https://doi.org/10.9771/cmbio.v23i3.57207
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of macrogeometry on primary stability and insertion torque in dental implants.

Palena Araújo Pinto, Meire Coelho Ferreira, Oswaldo Costa Neto, Rudys Rodolfo de Jesus Tavarez et al.
Revista de Ciências Médicas e Biológicas
Dental Implant Techniques and Outcomes
article

Influence of macrogeometry on primary stability and insertion torque in dental implants.

Palena Araújo Pinto, Meire Coelho Ferreira, Oswaldo Costa Neto, Rudys Rodolfo de Jesus Tavarez, Etevaldo Matos Maia Filho, Letícia Geovana Barros De Alencar
article en

Abstract

Introduction: The primary stability and insertion torque of osseintegrated implants is fundamental to the success of osseointegration, especially in cases where a prosthesis with immediate load is installed. Objetive: This study aimed to assess the influence of the macrogeometry of two implant types on insertion torque and primary stability. Materials and Methods: This study utilized 40 implants, comprising 20 conical and 20 cylindrical, divided into four experimental groups. These implants were installed by a single operator, in a synthetic bone block following the manufacturer's recommendations and the sub-instrumentation technique. Implant stability was immediately assessed post-installation using magnetic resonance frequency analysis, measured by Osstellâ. The insertion torque was determined using a surgical implant installation motor. The resulting data were subjected to statistical analysis using the Kruskal–Wallis test, with Dunn's post hoc test employed to determine any significant differences between the implant groups in terms of insertion torque and primary stability. Results: A significant difference was observed in the stability of buccolingual (p=0.002) and the average stability (p=0.008), with respective effect sizes of 0.197 and 0.335. The insertion torque also demonstrated significance (p=0.018). Notably, a decrease in the values of the 04×10 implants was observed when compared to the 3.5×10 implants. Conclusions: Implants with a smaller diameter exhibited greater primary stability, whereas those with a larger diameter demonstrated higher insertion torque values. Keywords: Dental implant; primary stability; insertion torque.

Revista de Ciências Médicas e BiológicasVol. 23(3)
Universidade Ceuma (BR)
Openalex Percentile: Top 99%
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Influence of macrogeometry on primary stability and insertion torque in dental implants. — Palena Araújo Pinto, Meire Coelho Ferreira, et al. · Revista de Ciências Médicas e Biológicas (2025) | TGRS Research Map | TGRS