A Trace C<sub>2</sub>H<sub>2</sub> Detection Based on Near‐Infrared Dual‐Comb Spectroscopy

ABSTRACT To achieve low‐concentration measurement using near‐infrared dual‐comb spectroscopy (DCS), this study presents a trace C 2 H 2 detection using near‐infrared DCS. A multi‐pass gas cell is employed to extend the optical path length, improving the sensor's sensitivity. Signal post‐processing via interpolation method is applied to reduce noise and achieve a flatter dual‐comb. The sensor's sensitivity and transient characteristics are evaluated by selecting absorption lines at 6526.53, 6529.17, and 6531.78 cm −1 , with the experimental profile being well‐fitted by a Voigt function. The stability of the system is confirmed with a C 2 H 2 concentration of 200 ppm over a 2.5 h observation period, showing a stability better than 2.305 × 10 −2 . Allan variance analysis indicates an optimal integration time of 597 s, yielding a minimum detection limit of 6.15 × 10 −4 for absorbance and 350 ppb for concentration. Additionally, the detection of nine absorption lines demonstrates the multi‐peak measurement capability, with deviations ranging from 0.059% to 12.496%.

Authors

Institutions

Publication Details

Journal
Microwave and Optical Technology Letters
Published
2025-01-01
DOI
https://doi.org/10.1002/mop.70073
Primary Topic
Spectroscopy and Laser Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Trace C2H2 Detection Based on Near‐Infrared Dual‐Comb Spectroscopy

Shaowen Rong, Yuqi Yang, Mengpeng Hu, Yueting Zhou et al.
Microwave and Optical Technology Letters
Spectroscopy and Laser Applications
article

A Trace C2H2 Detection Based on Near‐Infrared Dual‐Comb Spectroscopy

Shaowen Rong, Yuqi Yang, Mengpeng Hu, Yueting Zhou, Xiaocong Sun, Qiang Wang, Chuanliang Li, Ting Gong, Xuanbing Qiu
article en

Abstract

ABSTRACT To achieve low‐concentration measurement using near‐infrared dual‐comb spectroscopy (DCS), this study presents a trace C 2 H 2 detection using near‐infrared DCS. A multi‐pass gas cell is employed to extend the optical path length, improving the sensor's sensitivity. Signal post‐processing via interpolation method is applied to reduce noise and achieve a flatter dual‐comb. The sensor's sensitivity and transient characteristics are evaluated by selecting absorption lines at 6526.53, 6529.17, and 6531.78 cm −1 , with the experimental profile being well‐fitted by a Voigt function. The stability of the system is confirmed with a C 2 H 2 concentration of 200 ppm over a 2.5 h observation period, showing a stability better than 2.305 × 10 −2 . Allan variance analysis indicates an optimal integration time of 597 s, yielding a minimum detection limit of 6.15 × 10 −4 for absorbance and 350 ppb for concentration. Additionally, the detection of nine absorption lines demonstrates the multi‐peak measurement capability, with deviations ranging from 0.059% to 12.496%.

Microwave and Optical Technology LettersVol. 67(1)
Chinese Academy of Sciences (CN), Changchun Institute of Optics, Fine Mechanics and Physics (CN), Taiyuan University of Science and Technology (CN)
Openalex Percentile: Top 99%
Spectroscopy and Laser Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.