Calibration and Uncertainty Analysis of Scanning Electron Microscope Length Measurement Error
Based on the calibration method of scanning electron microscope length measurement error based on equidistant grid standard template, establish an evaluation model for measurement uncertainty and analyze the sources of various uncertainties. Based on measurement examples, this paper elaborates on the process of evaluating the measurement uncertainty of length indication error in scanning electron microscopy based on equidistant grid standard templates, calculates its extended uncertainty, and provides a reference for evaluating the measurement uncertainty of length indication error in scanning electron microscopy based on equidistant grid standard templates.
Authors
- Yi‐Ching Yang (ORCID: https://orcid.org/0000-0003-1391-8040)
- Jiajun Tang
- Tao Wang (ORCID: https://orcid.org/0000-0002-4294-9690)
- Yuzhou Chang (ORCID: https://orcid.org/0000-0003-4893-1886)
- Wei Zhang (ORCID: https://orcid.org/0000-0002-2601-3198)
Institutions
- Shanghai Institute of Measurement and Testing Technology (CN)
Publication Details
- Journal
- Highlights in Science Engineering and Technology
- Published
- 2025-01-10
- DOI
- https://doi.org/10.54097/pga99876
- Primary Topic
- Electron and X-Ray Spectroscopy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00