Design, Utilization, and Performance Enhancement of Instrumentation Amplifiers in Biomedical Applications
Instrumentation amplifiers (InAmps) play a crucial role in biomedical instrumentation by amplifying small differential signals while attenuating common-mode noise. This research paper focuses on the design utilization and performance optimization of InAmps for biomedical applications. Key aspects explored include noise reduction, common-mode rejection ratio (CMRR), and power efficiency. Simulations conducted using MATLAB Simulink demonstrate that chopper stabilization significantly reduces low-frequency noise, improving the signal-to-noise ratio (SNR) by approximately 15 dB. Utilizing low-noise operational amplifiers decreases input-referred noise by 30%, enhancing signal clarity. Precision resistor matching improves CMRR by 20 dB, while balanced layouts contribute an additional 10 dB improvement by minimizing electromagnetic interference. Operating amplifiers at lower supply voltages reduces power consumption by 40%, with a 25% reduction achieved through careful component selection. These findings offer valuable insights into designing robust and efficient instrumentation amplifiers for various biomedical applications, paving the way for advancements in portable and reliable health monitoring systems.
Authors
- Kamlesh Kumar Singh (ORCID: https://orcid.org/0000-0002-9382-6068)
- Ritik Singh
Institutions
- Amity University (AE)
Publication Details
- Published
- 2024-09-12
- DOI
- https://doi.org/10.1109/sparc61891.2024.10829174
- Primary Topic
- Analog and Mixed-Signal Circuit Design
- Type
- article
- Field-Weighted Citation Impact
- 0.00