A frequency-domain index for quickly selecting optimal down-sampling factor in photoacoustic imaging
Photoacoustic Microscopy (PAM) has emerged as a rapidly advancing non-invasive medical imaging method in recent years. However, slow imaging speed has impeded its widespread adoption in clinical applications. In certain scenarios, sparse spatial sampling is essential for PAM so there is a trade-off between spatial resolution and imaging speed. To address this limitation, we propose a frequency domain index based on Cumulative Power Difference (CPD) to determine rapidly the optimal down-sampling factors. In this study, the structural images of mice ears were acquired by the PAM system. Subsequently, the optimal down-sampling factor was determined through CPD analysis of these images via interpolation. Finally, the correlation between cumulative power difference and the image quality loss curve were analyzed. The results indicate that the quality of the reconstructed images decreases with the increasing down-sampling factor. Moreover, the cumulative power difference is an effective tool for rapid assessment of reconstruction image quality degradation due to varying down-sampling factors.
Authors
- Yameng Zhang (ORCID: https://orcid.org/0009-0008-8654-9892)
- Min Wan (ORCID: https://orcid.org/0009-0002-1581-5643)
- Jiani Li (ORCID: https://orcid.org/0000-0002-8258-3654)
- Weitao Li (ORCID: https://orcid.org/0000-0001-7855-428X)
- Ling Tao (ORCID: https://orcid.org/0000-0002-7076-1185)
- Shihao Tang
Institutions
- Nanjing Institute of Technology (CN)
- Nanjing University of Aeronautics and Astronautics (CN)
Publication Details
- Published
- 2025-01-10
- DOI
- https://doi.org/10.1117/12.3057116
- Primary Topic
- Photoacoustic and Ultrasonic Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00