Nyquist ghost elimination for diffusion MRI by dual-polarity readout at low b-values
Dual-polarity readout is a simple and robust way to mitigate Nyquist ghosting in diffusion-weighted echo-planar imaging but imposes doubled scan time. We here propose how dual-polarity readout can be implemented with little or no increase in scan time by exploiting an observed b-value dependence and signal averaging. The b-value dependence was confirmed in healthy volunteers with distinct ghosting at low b-values but of negligible magnitude at<i>b</i>= 1000 s/mm<sup>2</sup>. The usefulness of the suggested strategy was exemplified with a scan using tensor-valued diffusion encoding for estimation of parameter maps of mean diffusivity, and anisotropic and isotropic mean kurtosis, showing that ghosting propagated into all three parameter maps unless dual-polarity readout was applied. Results thus imply that extending the use of dual-polarity readout to low non-zero b-values provides effective ghost elimination and can be used without increased scan time for any diffusion MRI scan containing signal averaging at low b-values.
Authors
- Maria Ljungberg (ORCID: https://orcid.org/0000-0001-8991-2353)
- Isabella M. Björkman‐Burtscher (ORCID: https://orcid.org/0000-0002-9023-3363)
- Oscar Jalnefjord (ORCID: https://orcid.org/0000-0003-2741-5890)
- Nicolas Geades
- Guillaume Gilbert (ORCID: https://orcid.org/0000-0002-9960-1062)
Institutions
- Philips (Finland) (FI)
- Sahlgrenska University Hospital (SE)
- Region Västra Götaland (SE)
- CARE Canada (CA)
- Healthcare in Sweden (SE)
- Philips (Sweden) (SE)
- Philips (Canada) (CA)
- University of Gothenburg (SE)
Publication Details
- Journal
- Biomedical Physics & Engineering Express
- Published
- 2025-01-10
- DOI
- https://doi.org/10.1088/2057-1976/ada8b0
- Citations
- 1
- Primary Topic
- Advanced Neuroimaging Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 1.36
Funders
- Cancerfonden
- Stiftelsen Assar Gabrielssons Fond