Setting up a new, reliable and traceable current output specification for a Rotamass TI Coriolis mass flow meter by applying JCGM 100:2008
In this paper a traceable and reliable specification for the current output of a Rotamass TI Coriolis mass flow meter is derived based on experimental results and applying the general rules for evaluating and expressing uncertainty in measurement presented in JCGM 100:2008. In this way we showed that the analog current output of a Rotamass TI transmitter is a very reliable and precise method to monitor process quantities like the mass flow rate.
Authors
- P. Reinshaus
- B. Wild
Institutions
- Yokogawa Electric (Japan) (JP)
Publication Details
- Journal
- Measurement Sensors
- Published
- 2025-01-09
- DOI
- https://doi.org/10.1016/j.measen.2024.101576
- Primary Topic
- Flow Measurement and Analysis
- Type
- article
- Field-Weighted Citation Impact
- 0.00