Experimental Study of the Reduction of the Dose to the Radiographic Operator Using an Additional Rectangular Collimator and Detector Holder During Handheld Exposure with a Portable Intraoral X-Ray Device

Background: Use of a rectangular collimator and a detector holder reduces the patient’s dose in intraoral radiography; however, their quantitative effects on reducing the operator’s dose are unknown. Objectives: This study aimed to evaluate the reduction in the operator’s dose when using a rectangular collimator and a detector holder during handheld exposure with a portable intraoral X-ray device. Materials and Methods: Dose was measured using a cylindrical polymethylmethacrylate phantom and two semiconductor dosimeters. Stray radiation from the phantom and the X-ray device was measured at 15-degree intervals in horizontal and vertical circles with radii of 50 cm and 100 cm from the center of the phantom. Doses were compared with and without the additional rectangular collimator and the detector holder. Results: Use of the additional rectangular collimator and the detector holder reduced the operator’s dose to one-eighth and one-fifth, respectively, of the dose when these items were not used. Conclusion: Using an additional rectangular collimator and a detector holder appears effective for reducing the dose to the radiographic operator during handheld exposure with a portable intraoral X-ray device and could be applied to situations such as dental home health care and person identification in forensic dentistry.

Authors

Institutions

Publication Details

Journal
Journal of Indian Academy of Oral Medicine and Radiology
Published
2024-10-01
DOI
https://doi.org/10.4103/jiaomr.jiaomr_332_23
Citations
1
Primary Topic
Dental Radiography and Imaging
Type
article
Field-Weighted Citation Impact
0.49
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Experimental Study of the Reduction of the Dose to the Radiographic Operator Using an Additional Rectangular Collimator and Detector Holder During Handheld Exposure with a Portable Intraoral X-Ray Device

Noriyoshi Shiba, Nobutoshi Takahashi, Yusei Otaka, Yasuo Harata et al.
1 citations
Journal of Indian Academy of Oral Medicine and Radiology
Dental Radiography and Imaging
0.49
article

Experimental Study of the Reduction of the Dose to the Radiographic Operator Using an Additional Rectangular Collimator and Detector Holder During Handheld Exposure with a Portable Intraoral X-Ray Device

Noriyoshi Shiba, Nobutoshi Takahashi, Yusei Otaka, Yasuo Harata, Shinji Kito, Maki Izawa
article en
1 citations

Abstract

Background: Use of a rectangular collimator and a detector holder reduces the patient’s dose in intraoral radiography; however, their quantitative effects on reducing the operator’s dose are unknown. Objectives: This study aimed to evaluate the reduction in the operator’s dose when using a rectangular collimator and a detector holder during handheld exposure with a portable intraoral X-ray device. Materials and Methods: Dose was measured using a cylindrical polymethylmethacrylate phantom and two semiconductor dosimeters. Stray radiation from the phantom and the X-ray device was measured at 15-degree intervals in horizontal and vertical circles with radii of 50 cm and 100 cm from the center of the phantom. Doses were compared with and without the additional rectangular collimator and the detector holder. Results: Use of the additional rectangular collimator and the detector holder reduced the operator’s dose to one-eighth and one-fifth, respectively, of the dose when these items were not used. Conclusion: Using an additional rectangular collimator and a detector holder appears effective for reducing the dose to the radiographic operator during handheld exposure with a portable intraoral X-ray device and could be applied to situations such as dental home health care and person identification in forensic dentistry.

Journal of Indian Academy of Oral Medicine and RadiologyVol. 36(4)
Meikai University (JP)
Good health and well-being
Openalex Percentile: Top 31%
Dental Radiography and Imaging
0.49
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.