Test-Retest Reliability and Validity of TecnoBody D-Wall to Assess the Range of Motion During Overhead Squat in Healthy Individuals

This study evaluated the validity and reliability of the TecnoBody D-Wall system in assessing joint range of motion (ROM) during overhead squat movements in healthy individuals, using Kinovea as a reference tool for data comparison. A total of 29 participants (16 males, 13 females) with a mean age of 28.41 ± 6.66 years were included. Measurements were conducted for hip and knee joint angles in the sagittal plane, with three repetitions per participant analyzed using both systems. The D-Wall system employed a 3D Kinect V2 camera and force platform, while Kinovea used 2D video-based motion analysis. The results demonstrated excellent agreement between the two systems, with intra-class correlation coefficients (ICC) ranging from 0.79 to 0.99. For the knee joint, the test-retest ICC values were 0.99 (95% CI: 0.97-0.99) for Kinovea and 0.98 (95% CI: 0.95-0.99) for the D-Wall on the right side, and 0.98 (95% CI: 0.97-0.99) for Kinovea and 0.88 (95% CI: 0.79-0.94) for the D-Wall on the left side. For the hip joint, test-retest ICC values were 0.99 (95% CI: 0.97-0.99) for Kinovea and 0.94 (95% CI: 0.88-0.97) for the D-Wall on the right side, and 0.98 (95% CI: 0.97-0.99) for Kinovea and 0.93 (95% CI: 0.87-0.97) for the D-Wall on the left side. Bland-Altman plots confirmed good agreement, with no significant systematic bias observed. Both systems showed statistically insignificant differences (<i>p</i> > 0.05) between measurements, and correlation values ranged from 0.83 to 0.99, indicating strong associations. These findings highlight the high validity and reliability of both TecnoBody D-Wall and Kinovea systems in measuring joint angles during dynamic movements. The comparable accuracy between systems suggests that either system can be effectively utilized in clinical or research settings, depending on specific needs and resource availability.

Authors

Institutions

Publication Details

Journal
Life
Published
2025-01-10
DOI
https://doi.org/10.3390/life15010080
Citations
1
Primary Topic
Sports Performance and Training
Type
article
Field-Weighted Citation Impact
1.72
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Test-Retest Reliability and Validity of TecnoBody D-Wall to Assess the Range of Motion During Overhead Squat in Healthy Individuals

Türker Bıyıklı, Pervin Demır, Sinan Seyhan, Çağlar Soylu et al.
1 citations
Life
Sports Performance and Training
1.72
article

Test-Retest Reliability and Validity of TecnoBody D-Wall to Assess the Range of Motion During Overhead Squat in Healthy Individuals

Türker Bıyıklı, Pervin Demır, Sinan Seyhan, Çağlar Soylu, Görkem Açar, Berkay Üzümcü
article en
1 citations

Abstract

This study evaluated the validity and reliability of the TecnoBody D-Wall system in assessing joint range of motion (ROM) during overhead squat movements in healthy individuals, using Kinovea as a reference tool for data comparison. A total of 29 participants (16 males, 13 females) with a mean age of 28.41 ± 6.66 years were included. Measurements were conducted for hip and knee joint angles in the sagittal plane, with three repetitions per participant analyzed using both systems. The D-Wall system employed a 3D Kinect V2 camera and force platform, while Kinovea used 2D video-based motion analysis. The results demonstrated excellent agreement between the two systems, with intra-class correlation coefficients (ICC) ranging from 0.79 to 0.99. For the knee joint, the test-retest ICC values were 0.99 (95% CI: 0.97-0.99) for Kinovea and 0.98 (95% CI: 0.95-0.99) for the D-Wall on the right side, and 0.98 (95% CI: 0.97-0.99) for Kinovea and 0.88 (95% CI: 0.79-0.94) for the D-Wall on the left side. For the hip joint, test-retest ICC values were 0.99 (95% CI: 0.97-0.99) for Kinovea and 0.94 (95% CI: 0.88-0.97) for the D-Wall on the right side, and 0.98 (95% CI: 0.97-0.99) for Kinovea and 0.93 (95% CI: 0.87-0.97) for the D-Wall on the left side. Bland-Altman plots confirmed good agreement, with no significant systematic bias observed. Both systems showed statistically insignificant differences (<i>p</i> > 0.05) between measurements, and correlation values ranged from 0.83 to 0.99, indicating strong associations. These findings highlight the high validity and reliability of both TecnoBody D-Wall and Kinovea systems in measuring joint angles during dynamic movements. The comparable accuracy between systems suggests that either system can be effectively utilized in clinical or research settings, depending on specific needs and resource availability.

LifeVol. 15(1)
Manisa Celal Bayar University (TR), Sağlık Bilimleri Üniversitesi (TR), Ankara Yıldırım Beyazıt University (TR), Robotic Technology (United States) (US), Marmara University (TR)
Openalex Percentile: Top 24%
Sports Performance and Training
1.72
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.