Development of a prototype ChLIA and a research‐use only ELISA for the detection of light chain neurofilament (NfL) in CSF

Abstract Background Increased levels of the light chain of neurofilament (NfL) in human cerebrospinal fluid indicate axonal damage. Their determination can support laboratory diagnosis and monitoring of neurological diseases associated with axonal damage. We developed a prototype chemiluminescence assay (ChLIA) and an enzyme‐linked immunoassay (ELISA, research use only) to detect NfL and report their analytical agreement with an established ELISA. Method Agreement between the established NF‐light ELISA (UMANDiagnostics) and the Neurofilament Light (NfL) ELISA (EUROIMMUN) as well as the prototype NfL ChLIA (EUROIMMUN) was determined using 19 internal reference samples containing native NfL, with NfL levels distributed across the measurement range of the NF‐light ELISA. Possible bias between methods was estimated with Passing‐Bablok regression and Kolmogorov‐Smirnov CUSUM test. Result Comparison of NF‐light ELISA and Neurofilament Light (NfL) ELISA indicated a small proportional deviation with NF‐light ELISA values being higher than Neurofilament Light (NfL) ELISA values (regression: y = ‐63.9+0.97x, 95% CI intercept: [‐490.8, ‐79.67], CI slope: [0.89, 1.05], r = 0.08, CUSUM test p = 0.6). Comparison of Neurofilament Light (NfL) ELISA and prototype NfL ChLIA indicated a small shift with ChLIA values being higher than ELISA values, especially in the upper end of the measurement range (regression: y = 13.73+1.14x, CI intercept: [‐140.9, 97.05], CI slope: [1.07, 1.23], r = 0.98, CUSUM test p = 0.6). Conclusion The results of analytical agreement between NfL levels determined by the ELISA and the prototype ChLIA in comparison to the established ELISA show promise regarding their performance characteristics. Being limited by the small sample size, these results will be followed up by a larger method comparison and validation study.

Authors

Institutions

Publication Details

Journal
Alzheimer s & Dementia
Published
2024-12-01
DOI
https://doi.org/10.1002/alz.094720
Primary Topic
Cerebrospinal fluid and hydrocephalus
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a prototype ChLIA and a research‐use only ELISA for the detection of light chain neurofilament (NfL) in CSF

Victor Herbst, Viola Borchardt, Katharina Zeplin, Stephanie Franco et al.
Alzheimer s & Dementia
Cerebrospinal fluid and hydrocephalus
article

Development of a prototype ChLIA and a research‐use only ELISA for the detection of light chain neurofilament (NfL) in CSF

Victor Herbst, Viola Borchardt, Katharina Zeplin, Stephanie Franco, Julia Niewand
article en

Abstract

Abstract Background Increased levels of the light chain of neurofilament (NfL) in human cerebrospinal fluid indicate axonal damage. Their determination can support laboratory diagnosis and monitoring of neurological diseases associated with axonal damage. We developed a prototype chemiluminescence assay (ChLIA) and an enzyme‐linked immunoassay (ELISA, research use only) to detect NfL and report their analytical agreement with an established ELISA. Method Agreement between the established NF‐light ELISA (UMANDiagnostics) and the Neurofilament Light (NfL) ELISA (EUROIMMUN) as well as the prototype NfL ChLIA (EUROIMMUN) was determined using 19 internal reference samples containing native NfL, with NfL levels distributed across the measurement range of the NF‐light ELISA. Possible bias between methods was estimated with Passing‐Bablok regression and Kolmogorov‐Smirnov CUSUM test. Result Comparison of NF‐light ELISA and Neurofilament Light (NfL) ELISA indicated a small proportional deviation with NF‐light ELISA values being higher than Neurofilament Light (NfL) ELISA values (regression: y = ‐63.9+0.97x, 95% CI intercept: [‐490.8, ‐79.67], CI slope: [0.89, 1.05], r = 0.08, CUSUM test p = 0.6). Comparison of Neurofilament Light (NfL) ELISA and prototype NfL ChLIA indicated a small shift with ChLIA values being higher than ELISA values, especially in the upper end of the measurement range (regression: y = 13.73+1.14x, CI intercept: [‐140.9, 97.05], CI slope: [1.07, 1.23], r = 0.98, CUSUM test p = 0.6). Conclusion The results of analytical agreement between NfL levels determined by the ELISA and the prototype ChLIA in comparison to the established ELISA show promise regarding their performance characteristics. Being limited by the small sample size, these results will be followed up by a larger method comparison and validation study.

Alzheimer s & DementiaVol. 20(S8)
Euroimmun Medizinische Labordiagnostika (Germany) (DE)
Openalex Percentile: Top 73%
Cerebrospinal fluid and hydrocephalus
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Development of a prototype ChLIA and a research‐use only ELISA for the detection of light chain neurofilament (NfL) in CSF — Victor Herbst, Viola Borchardt, et al. · Alzheimer s & Dementia (2024) | TGRS Research Map | TGRS