Simulation of high signal-to-noise ratio resonant photodetector for homodyne measurement and its verification
In this paper, two models for simulating the shot noise and electronic noise performances of resonant photodetectors designed for homodyne measurements are presented. One is based on a combination of a buffer and a low-noise amplifier, and the other is based on an operational amplifier. Through the comparisons between the numerical simulation results and the experimentally obtained data, excellent agreements are achieved, which show that the models provide a highly efficient guide for the development of a high signal-to-noise ratio (SNR) resonant photodetector. Furthermore, we demonstrate a high SNR resonant photodetector for homodyne measurements at the 147 MHz optical sideband, achieving a 20.8 dB SNR of the shot noise to the electronic noise with a 2 mW optical signal input, utilizing a combination of a buffer and a low-noise amplifier. Concurrently, we have obtained another resonant photodetector at the 1.14 GHz optical sideband, which exhibits a 13 dB SNR based on an operational amplifier.
Authors
- S. B. Wang
- Ruifang Dong (ORCID: https://orcid.org/0000-0001-6706-8980)
- Guangqian Wu
- Wei Liu (ORCID: https://orcid.org/0000-0003-2655-512X)
- Xuyin Niu
Institutions
- Shanxi University (CN)
- Chinese Academy of Sciences (CN)
- National Time Service Center (CN)
- Astronomy and Space (AU)
- University of Chinese Academy of Sciences (CN)
Publication Details
- Journal
- Review of Scientific Instruments
- Published
- 2025-01-01
- DOI
- https://doi.org/10.1063/5.0235927
- Citations
- 1
- Primary Topic
- Photonic and Optical Devices
- Type
- article
- Field-Weighted Citation Impact
- 0.73
Funders
- National Natural Science Foundation of China
- China Postdoctoral Science Foundation
- Youth Innovation Promotion Association of the Chinese Academy of Sciences