Commutability assessment of new standard reference materials (SRMs) for determining serum total 25-hydroxyvitamin D using ligand binding and liquid chromatography–tandem mass spectrometry (LC–MS/MS) assays
Commutability is where the measurement response for a reference material (RM) is the same as for an individual patient sample with the same concentration of analyte measured using two or more measurement systems. Assessment of commutability is essential when the RM is used in a calibration hierarchy or to ensure that clinical measurements are comparable across different measurement procedures and at different times. The commutability of three new Standard Reference Materials<sup>®</sup> (SRMs) for determining serum total 25-hydroxyvitamin D [25(OH)D], defined as the sum of 25-hydroxyvitamin D<sub>2</sub> [25(OH)D<sub>2</sub>] and 25-hydroxyvitamin D<sub>3</sub> [25(OH)D<sub>3</sub>], was assessed through an interlaboratory study. The following SRMs were assessed: (1) SRM 2969 Vitamin D Metabolites in Frozen Human Serum (Total 25-Hydroxyvitamin D Low Level), (2) SRM 2970 Vitamin D Metabolites in Frozen Human Serum (25-Hydroxyvitamin D<sub>2</sub> High Level), and (3) SRM 1949 Frozen Human Prenatal Serum. These SRMs represent three clinically relevant situations including (1) low levels of total 25(OH)D, (2) high level of 25(OH)D<sub>2</sub>, and (3) 25(OH)D levels in nonpregnant women and women during each of the three trimesters of pregnancy with changing concentrations of vitamin D-binding protein (VDBP). Twelve laboratories using 17 different ligand binding assays and eight laboratories using nine commercial and custom liquid chromatography-tandem mass spectrometry (LC-MS/MS) assays provided results in this study. Commutability of the SRMs with patient samples was assessed using the Clinical and Laboratory Standards Institute (CLSI) approach based on 95% prediction intervals or a pre-set commutability criterion and the recently introduced International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) approach based on differences in bias for the clinical and reference material samples using a commutability criterion of 8.8%. All three SRMs were deemed as commutable with all LC-MS/MS assays using both CLSI and IFCC approaches. SRM 2969 and SRM 2970 were deemed noncommutable for three and seven different ligand binding assays, respectively, when using the IFCC approach. Except for two assays, one or more of the three pregnancy levels of SRM 1949 were deemed noncommutable or inconclusive using different ligand binding assays and the commutability criterion of 8.8%. Overall, a noncommutable assessment for ligand binding assays is determined for these SRMs primarily due to a lack of assay selectivity related to 25(OH)D<sub>2</sub> or an increasing VDBP in pregnancy trimester materials rather than the quality of the SRMs. With results from 17 different ligand binding and nine LC-MS/MS assays, this study provides valuable knowledge for clinical laboratories to inform SRM selection when assessing 25(OH)D status in patient populations, particularly in subpopulations with low levels of 25(OH)D, high levels of 25(OH)D<sub>2</sub>, women only, or women who are pregnant.
Authors
- Emma L. Williams (ORCID: https://orcid.org/0000-0002-0309-129X)
- Hisao Tsukamoto (ORCID: https://orcid.org/0000-0003-4138-7398)
- Adam J. Kuszak (ORCID: https://orcid.org/0000-0001-8719-145X)
- Chao Dou (ORCID: https://orcid.org/0000-0003-0626-240X)
- Grace Hahm
- Camille Pease
- Pierre Lukas (ORCID: https://orcid.org/0000-0002-1071-9057)
- Ben Saida
- Jennifer Cheek
- Satoshi Kojima
- Stephen A. Wise (ORCID: https://orcid.org/0000-0002-0021-1970)
- Sun Jong You
- Andrew N. Hoofnagle (ORCID: https://orcid.org/0000-0002-6449-0243)
- Neil H. Parker
- Carole Tourneur
- Jan Schultess
- Christian Vogl
- Ekaterina Mineva (ORCID: https://orcid.org/0000-0002-9079-0856)
- Christine M Pfeiffer (ORCID: https://orcid.org/0000-0003-3438-5191)
- Hope A. Weiler
- James Freeman (ORCID: https://orcid.org/0000-0002-0429-1322)
- Kevin D. Cashman (ORCID: https://orcid.org/0000-0002-8185-1039)
- Stéphanie Peeters
- Karen Galvin
- Heather Pham
- Ralf Fischer
- Étienne Cavalier (ORCID: https://orcid.org/0000-0003-0947-2226)
- Alexandra Bennett (ORCID: https://orcid.org/0000-0002-5403-0382)
- A Bielecki
- Caroline Le Goff (ORCID: https://orcid.org/0000-0001-7816-3705)
- Hubert W. Vesper (ORCID: https://orcid.org/0000-0002-9649-9015)
- Carolyn Q Burdette (ORCID: https://orcid.org/0000-0002-0843-9224)
- Bin Wei (ORCID: https://orcid.org/0000-0001-5076-4082)
- Johanna E Camara (ORCID: https://orcid.org/0000-0002-9415-8452)
- Ashley S.P. Boggs (ORCID: https://orcid.org/0000-0001-9765-7559)
- Federica Nalin
- Christian Beckert
- Jinghong Li (ORCID: https://orcid.org/0000-0002-0750-7352)
- Samantha Pagliaro
- Kevin Wang (ORCID: https://orcid.org/0000-0002-9343-6473)
- Christian Popp (ORCID: https://orcid.org/0000-0001-6924-313X)
- Amit Ghoshal
- Hsuan-Chieh Liao
- Dominik Osterritter
- Jeffery R. Budd
- Laura E. Briggs
Institutions
- Health Canada (CA)
- Centers for Disease Control and Prevention (US)
- National Institutes of Health (US)
- National Institute of Standards and Technology (US)
- Imperial College Healthcare NHS Trust (GB)
- University of Liège (BE)
- University of Washington (US)
- University College Cork (IE)
- Roche Pharma AG (Germany) (DE)
- Ingegneria dei Sistemi (United Kingdom) (GB)
- Office of Dietary Supplements (US)
- Abbott (Germany) (DE)
- TSI (United States) (US)
- Fujirebio (Japan) (JP)
- Centre Hospitalier Universitaire de Liège (BE)
- Diazyme (United States) (US)
- bioMérieux (France) (FR)
- Siemens Healthcare (United States) (US)
- PerkinElmer (United States) (US)
- Tosoh (Japan) (JP)
- Quest Diagnostics (United States) (US)
Publication Details
- Journal
- Analytical and Bioanalytical Chemistry
- Published
- 2025-01-10
- DOI
- https://doi.org/10.1007/s00216-024-05699-7
- Citations
- 6
- Primary Topic
- Vitamin D Research Studies
- Type
- article
- Field-Weighted Citation Impact
- 8.83
Funders
- Office of Dietary Supplements
- National Institutes of Health
- National Institute of Standards and Technology