Temperature Dependence of the Sensitivity of Hydrophones for Biomedical Ultrasound Exposimetry

Though it has been recognized that the sensitivity of hydrophones depends on the temperature of the water they are used in, the amount of specific data that is available is very limited. This is possible because the measurements are technically challenging, laborious, and time-consuming. A broadband primary hydrophone calibration setup was extended to implement stable calibration conditions at different temperature settings. Four hydrophones of different types commonly used in ultrasound exposimetry were then characterized in the range from 17 °C to 29 °C, and average change rates of the sensitivity with temperature were determined for different ultrasonic frequencies. Two different membrane hydrophones showed an increase in sensitivity with increasing temperature in the range from 0.55% to 1.10% per 1 °C temperature rise within their bandwidths. The results for a capsule-type and a needle-type hydrophone were different in the sense that a decreasing sensitivity with increasing temperature was also observed. For the capsule-type hydrophone, a small increase was observed up to 15 MHz and a decrease for higher frequencies. The needle-type hydrophone provided a decrease at all frequencies, and the results were more noisy. Data, as determined in this study, can be applied to correct acoustic output measurements of medical ultrasonic equipment if the water temperature of the hydrophone application differs from that during calibration. Alternatively, it may suffice in some applications to consider a sensitivity change with temperature within the uncertainty estimation of the exposure measurement, in particular, if the temperature difference is limited to 1 °C or 2 °C.

Authors

Institutions

Publication Details

Journal
IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control
Published
2025-01-09
DOI
https://doi.org/10.1109/tuffc.2025.3527625
Citations
2
Primary Topic
Hemodynamic Monitoring and Therapy
Type
article
Field-Weighted Citation Impact
4.01
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Temperature Dependence of the Sensitivity of Hydrophones for Biomedical Ultrasound Exposimetry

Volker Wilkens, Martin Weber, Jennifer Twiefel
2 citations
IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control
Hemodynamic Monitoring and Therapy
4.01
article

Temperature Dependence of the Sensitivity of Hydrophones for Biomedical Ultrasound Exposimetry

Volker Wilkens, Martin Weber, Jennifer Twiefel
article en
2 citations

Abstract

Though it has been recognized that the sensitivity of hydrophones depends on the temperature of the water they are used in, the amount of specific data that is available is very limited. This is possible because the measurements are technically challenging, laborious, and time-consuming. A broadband primary hydrophone calibration setup was extended to implement stable calibration conditions at different temperature settings. Four hydrophones of different types commonly used in ultrasound exposimetry were then characterized in the range from 17 °C to 29 °C, and average change rates of the sensitivity with temperature were determined for different ultrasonic frequencies. Two different membrane hydrophones showed an increase in sensitivity with increasing temperature in the range from 0.55% to 1.10% per 1 °C temperature rise within their bandwidths. The results for a capsule-type and a needle-type hydrophone were different in the sense that a decreasing sensitivity with increasing temperature was also observed. For the capsule-type hydrophone, a small increase was observed up to 15 MHz and a decrease for higher frequencies. The needle-type hydrophone provided a decrease at all frequencies, and the results were more noisy. Data, as determined in this study, can be applied to correct acoustic output measurements of medical ultrasonic equipment if the water temperature of the hydrophone application differs from that during calibration. Alternatively, it may suffice in some applications to consider a sensitivity change with temperature within the uncertainty estimation of the exposure measurement, in particular, if the temperature difference is limited to 1 °C or 2 °C.

IEEE Transactions on Ultrasonics Ferroelectrics and Frequency ControlVol. 72(3)
Physikalisch-Technische Bundesanstalt (DE), University of Helsinki (FI)
Affordable and clean energy
Openalex Percentile: Top 9%
Hemodynamic Monitoring and Therapy
4.01
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.