Effect of Scanner on beta amyloid quantification in a head‐to‐head comparison of 18F‐NAV4694 Centiloid measurements between Biograph Vision, Gemini and Biograph mCT
Abstract Background New generation PET scanners achieve superior resolution and sensitivity, but the implications on beta amyloid (Aβ) quantitation are not well understood. The Centiloid (CL) scale (Klunk et al., 2015) was introduced to promote consistent Aβ burden quantification across different positron emission tomography (PET) tracers and quantification pipelines, but was not intended to control for hardware or reconstruction changes. Instead, post‐reconstruction smoothing harmonisation (Joshi, Koeppe, Fessler, 2009) is assumed to harmonise inter‐scanner Centiloid measurements. In this work, we compare paired Centiloid measurements between the Siemens Biograph Vision and one of Siemens Biograph mCT or Philips Gemini. Method 28 AIBL participants underwent MPRAGE‐MRI and 18 F‐NAV4694 imaging on two scanners (of Vision, mCT or Gemini) within a year, reconstructed per the ADNI protocols for each scanner. Centiloid was quantified using the standard SPM framework and corrected for expected accumulation between scans. We investigated difference in Centiloid and effectiveness of PET resolution harmonisation, i.e., smoothed to 6mm‐equivalent point spread function using estimates from Hoffman scans. Statistical significance was evaluated using empirical, one‐sided, non‐parametric bootstrap estimates. Results Figure 1 shows representative differences between the Vision and Gemini, and Figures 2,3 depict per‐scanner scatter and Bland‐Altman plots of Centiloid with and without resolution harmonisation, respectively. Even with harmonisation, Centiloid from Vision was found to be significantly higher than from Gemini (p=0.000) and from mCT (p=0.031); the difference followed a linear association with slope>1 (p=0.000 and p=0.010, respectively). The difference at 0CL was negligible, but at 100CL (Gemini) was +10.5% and at 100CL (mCT) was +11.7%. Resolution harmonisation only reduced ∼20% of scanner differences at most. Conclusion Scanner changes led to different Centiloid quantification, with difference proportional to the participant’s Centiloid itself. The standard resolution harmonisation procedure did not correct for scanner difference. For high‐amyloid individuals, the measured difference between Vision and Gemini or mCT (> +10%) was greater than the expected annual accumulation rate and so significant for longitudinal studies. Some scanners may exaggerate measured changes in Centiloid, or longitudinal scanner changes may mask or amplify changes. The fact that the effect between scanners was found linear is promising that scanner calibration methods are possible, further investigation is warranted.
Authors
- Vincent Doré (ORCID: https://orcid.org/0000-0002-8051-0558)
- Svetlana Bozinovski
- Jürgen Fripp (ORCID: https://orcid.org/0000-0001-9705-0079)
- Graeme O’Keefe
- Ashley Gillman (ORCID: https://orcid.org/0000-0001-9130-1092)
- Shenpeng Li (ORCID: https://orcid.org/0000-0003-3021-3026)
- Pierrick Bourgeat (ORCID: https://orcid.org/0000-0002-2605-4766)
- Azadeh Feizpour
- Victor L. Villemagne (ORCID: https://orcid.org/0000-0002-5832-9875)
- Timothy Cox (ORCID: https://orcid.org/0000-0002-7232-5096)
- Robert Williams (ORCID: https://orcid.org/0000-0001-6060-5042)
- Natasha Krishnadas (ORCID: https://orcid.org/0000-0002-5374-2839)
- Kun Huang (ORCID: https://orcid.org/0000-0001-7378-5660)
- Rosita Shishegar (ORCID: https://orcid.org/0000-0003-4636-8145)
- Christopher C. Rowe
Institutions
- Commonwealth Scientific and Industrial Research Organisation (AU)
- The University of Melbourne (AU)
- University of Pittsburgh (US)
- Florey Institute of Neuroscience and Mental Health (AU)
- National Imaging Facility (AU)
- Australian e-Health Research Centre (AU)
- Austin Health (AU)
Publication Details
- Journal
- Alzheimer s & Dementia
- Published
- 2024-12-01
- DOI
- https://doi.org/10.1002/alz.087430
- Citations
- 2
- Primary Topic
- Medical Imaging Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.81