Quantitative determination of leptin hormone using gold nanoparticle-based lateral flow assay

A lateral flow assay (LFA) has been developed that can be used in point-of-care (PoC) use for the sensitive determination of leptin hormone. The limit of detection value was 0.158 ng/mL and the limit of quantification value was 0.479 ng/mL for leptin hormone. The stability studies showed that the assay response was equal to the initial value even after 6 months. LFA accuracy tests performed in commercial artificial serum samples were compared with the enzyme-linked immunosorbent assay (ELISA). The obtained recovery values between 95 and 110% show that the developed LFA can be used for quantitative determination of leptin hormone. As the result of the study, a fast and disposable assay was developed for the determination of leptin hormone, which can be performed with a small amount of sample in less than 15 min without the need for long-term analysis stages, qualified personnel, expensive equipment and devices.

Authors

Institutions

Publication Details

Journal
Microchimica Acta
Published
2025-01-09
DOI
https://doi.org/10.1007/s00604-024-06945-6
Citations
2
Primary Topic
Biosensors and Analytical Detection
Type
article
Field-Weighted Citation Impact
1.35

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quantitative determination of leptin hormone using gold nanoparticle-based lateral flow assay

Erhan Zor, Haluk Bingöl, Sabri Alpaydin
2 citations
Microchimica Acta
Biosensors and Analytical Detection
1.35
article

Quantitative determination of leptin hormone using gold nanoparticle-based lateral flow assay

Erhan Zor, Haluk Bingöl, Sabri Alpaydin
article en
2 citations

Abstract

A lateral flow assay (LFA) has been developed that can be used in point-of-care (PoC) use for the sensitive determination of leptin hormone. The limit of detection value was 0.158 ng/mL and the limit of quantification value was 0.479 ng/mL for leptin hormone. The stability studies showed that the assay response was equal to the initial value even after 6 months. LFA accuracy tests performed in commercial artificial serum samples were compared with the enzyme-linked immunosorbent assay (ELISA). The obtained recovery values between 95 and 110% show that the developed LFA can be used for quantitative determination of leptin hormone. As the result of the study, a fast and disposable assay was developed for the determination of leptin hormone, which can be performed with a small amount of sample in less than 15 min without the need for long-term analysis stages, qualified personnel, expensive equipment and devices.

Microchimica ActaVol. 192(2)
Necmettin Erbakan University (TR)
Türkiye Bilimsel ve Teknolojik Araştırma Kurumu, Necmettin Erbakan Üniversitesi
Openalex Percentile: Top 23%
Biosensors and Analytical Detection
1.35
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Quantitative determination of leptin hormone using gold nanoparticle-based lateral flow assay — Erhan Zor, Haluk Bingöl, et al. · Microchimica Acta (2025) | TGRS Research Map | TGRS