Changes in scanning orientation effects of Gafchromic EBT-3 film irradiated with ultra-high dose rate proton beams

While the FLASH radiotherapy (FLASH RT) using ultra-high-dose-rate (UHDR) proton beams has the potential to achieve more effective cancer treatment, reliable and practical dosimetry of UHDR beams remains challenging. Although the commercial radiochromic films, named “Gafchromic film,” can be promising tools for the measurements of UHDR beam profiles, these films have some unfavorable features such as scanning orientation effects and lateral response artifacts. In this study, we examined the scanning orientation effects of Gafchromic EBT3 films irradiated with UHDR proton beams. The square sheets (5.5 × 5.5 cm2) of EBT3 film were irradiated with 8.7–8.8 Gy(H2O) at the plateau region of 139 MeV proton beams in the treatment room at the Nagoya Proton Therapy Center (NPTC) at three different dose rates: 0.29 Gy s−1, 170 Gy s−1, and 720 Gy s−1. The irradiated films were scanned in the reference and perpendicular orientations using two scanners: a flatbed scanner (FBS) (GT-X980, Seiko Epson Corp.) and an overhead scanner (OHS) (Aura, CZUR Tech Co., Ltd.). At the conventional dose rate (0.29 Gy s−1), the inverted red and green color profiles acquired by FBS in the perpendicular orientation were significantly lower than those in the reference orientation, as observed in previous studies using photon beams. Unexpectedly, such orientation effects were not observed in the FBS images of the EBT3 films irradiated with UHDR beams. The scanning orientation did not affect the color profiles of the OHS images at any dose rate tested. These results imply that the scanning orientation effect of the Gafchromic film is not a concern in beam profile measurements for FLASH RT. More comprehensive studies are needed to explain the mechanisms underlying these phenomena and to confirm the universality of the findings obtained in this study.

Authors

Institutions

Publication Details

Journal
Discover Applied Sciences
Published
2025-01-09
DOI
https://doi.org/10.1007/s42452-025-06484-6
Citations
4
Primary Topic
Radiation Effects in Electronics
Type
article
Field-Weighted Citation Impact
2.92
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Changes in scanning orientation effects of Gafchromic EBT-3 film irradiated with ultra-high dose rate proton beams

Katsunori Yogo, Hiroshi Yasuda, Masashi Yamada, Masumi Umezawa et al.
4 citations
Discover Applied Sciences
Radiation Effects in Electronics
2.92
article

Changes in scanning orientation effects of Gafchromic EBT-3 film irradiated with ultra-high dose rate proton beams

Katsunori Yogo, Hiroshi Yasuda, Masashi Yamada, Masumi Umezawa, Hassna Bantan, T. Toshito
article en
4 citations

Abstract

While the FLASH radiotherapy (FLASH RT) using ultra-high-dose-rate (UHDR) proton beams has the potential to achieve more effective cancer treatment, reliable and practical dosimetry of UHDR beams remains challenging. Although the commercial radiochromic films, named “Gafchromic film,” can be promising tools for the measurements of UHDR beam profiles, these films have some unfavorable features such as scanning orientation effects and lateral response artifacts. In this study, we examined the scanning orientation effects of Gafchromic EBT3 films irradiated with UHDR proton beams. The square sheets (5.5 × 5.5 cm2) of EBT3 film were irradiated with 8.7–8.8 Gy(H2O) at the plateau region of 139 MeV proton beams in the treatment room at the Nagoya Proton Therapy Center (NPTC) at three different dose rates: 0.29 Gy s−1, 170 Gy s−1, and 720 Gy s−1. The irradiated films were scanned in the reference and perpendicular orientations using two scanners: a flatbed scanner (FBS) (GT-X980, Seiko Epson Corp.) and an overhead scanner (OHS) (Aura, CZUR Tech Co., Ltd.). At the conventional dose rate (0.29 Gy s−1), the inverted red and green color profiles acquired by FBS in the perpendicular orientation were significantly lower than those in the reference orientation, as observed in previous studies using photon beams. Unexpectedly, such orientation effects were not observed in the FBS images of the EBT3 films irradiated with UHDR beams. The scanning orientation did not affect the color profiles of the OHS images at any dose rate tested. These results imply that the scanning orientation effect of the Gafchromic film is not a concern in beam profile measurements for FLASH RT. More comprehensive studies are needed to explain the mechanisms underlying these phenomena and to confirm the universality of the findings obtained in this study.

Discover Applied SciencesVol. 7(1)
Hiroshima University (JP), Nagoya City University (JP), Nagoya University (JP), Hitachi (Japan) (JP)
Affordable and clean energy
Openalex Percentile: Top 10%
Radiation Effects in Electronics
2.92
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.