Quality Control and Optimization of Computed Tomography Dose Index Volume (CTDI<sub>vol</sub>) of LightSpeed RT16 Xtra CT Scanner

Dose estimation and quality control in computed tomography (CT) scanners are useful in controlling the dose of radiation given to patients while tests are carried out. The study was performed in a 16-slice Computed Tomography (CT) system of LightSpeed RT16 Xtra CT scanner. Quality control was done using a vendor-provided QA Phantom, and the six aspects of image quality were measured. For CT dosimetry, Computed Tomography Dose index volume (CTDIvol) was performed using Computed Tomography Dose Index (CTDI) Phantom. CTDI Phantom consists of three parts: Pediatric Head, Adult Head, and Adult Body Phantom. A 10 cm long pencil ion chamber DCT-10 was used to measure the dose at different positions inside the CTDI Phantom. Data were collected using MagicMax Universal software. For dose estimation of the CTDIvol Report of AAPM Task Group, 96 and 111 formalisms were used. For Pediatric Head, Adult Head, and Adult Body Phantom the measured CIDIvol was 61.04 mGy, 48.11 mGy, and 18.08 mGy respectively. The study has shown deviations of 7%, 15%, and 19% between estimated and console-displayed doses for Pediatric Head, Adult Head, and Adult Body scan techniques respectively. The six aspects of image quality measured by QA Phantom were found to be compatible with the specifications of the machine and CTDIvol measured by CTDI Phantom were found within a tolerance limit of ±20%. Hence, the QC and dosimetry of the mentioned machine are within the limit.

Authors

Institutions

Publication Details

Journal
International Journal of Medical Physics Clinical Engineering and Radiation Oncology
Published
2025-01-01
DOI
https://doi.org/10.4236/ijmpcero.2025.141001
Citations
3
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
4.08
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quality Control and Optimization of Computed Tomography Dose Index Volume (CTDIvol) of LightSpeed RT16 Xtra CT Scanner

Md. Anwarul Islam, Afroza Shelley, Rajada Khatun, Shirin Akter et al.
3 citations
International Journal of Medical Physics Clinical Engineering and Radiation Oncology
Radiation Dose and Imaging
4.08
article

Quality Control and Optimization of Computed Tomography Dose Index Volume (CTDIvol) of LightSpeed RT16 Xtra CT Scanner

Md. Anwarul Islam, Afroza Shelley, Rajada Khatun, Shirin Akter, Ashrafun Nahar Monika, Laila Shamima Sharmin, Umme Sadia Binte Kashem
article en
3 citations

Abstract

Dose estimation and quality control in computed tomography (CT) scanners are useful in controlling the dose of radiation given to patients while tests are carried out. The study was performed in a 16-slice Computed Tomography (CT) system of LightSpeed RT16 Xtra CT scanner. Quality control was done using a vendor-provided QA Phantom, and the six aspects of image quality were measured. For CT dosimetry, Computed Tomography Dose index volume (CTDIvol) was performed using Computed Tomography Dose Index (CTDI) Phantom. CTDI Phantom consists of three parts: Pediatric Head, Adult Head, and Adult Body Phantom. A 10 cm long pencil ion chamber DCT-10 was used to measure the dose at different positions inside the CTDI Phantom. Data were collected using MagicMax Universal software. For dose estimation of the CTDIvol Report of AAPM Task Group, 96 and 111 formalisms were used. For Pediatric Head, Adult Head, and Adult Body Phantom the measured CIDIvol was 61.04 mGy, 48.11 mGy, and 18.08 mGy respectively. The study has shown deviations of 7%, 15%, and 19% between estimated and console-displayed doses for Pediatric Head, Adult Head, and Adult Body scan techniques respectively. The six aspects of image quality measured by QA Phantom were found to be compatible with the specifications of the machine and CTDIvol measured by CTDI Phantom were found within a tolerance limit of ±20%. Hence, the QC and dosimetry of the mentioned machine are within the limit.

International Journal of Medical Physics Clinical Engineering and Radiation OncologyVol. 14(01)
University of Dhaka (BD), Square Hospitals (BD), Bangladesh Atomic Energy Commission (BD)
Openalex Percentile: Top 8%
Radiation Dose and Imaging
4.08
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Quality Control and Optimization of Computed Tomography Dose Index Volume (CTDI<sub>vol</sub>) of LightSpeed RT16 Xtra CT Scanner — Md. Anwarul Islam, Afroza Shelley, et al. · International Journal of Medical Physics Clinical Engineering and Radiation Oncology (2025) | TGRS Research Map | TGRS