Accident Prevention “Awareness” System for Intravenous Drip Infusion Operations - Construction and Evaluation

The purpose of this study was to evaluate an infusion rate monitoring system (Tritech, Inc.) in a clinical setting and to identify challenges in using the system as an alternative to an infusion pump to prevent inappropriate infusion rates (excessive or insufficient). The system monitors the infusion rate and sends an alert to the nurse call system if the rate deviates from a preset limit. Unlike infusion pumps, this system does not adjust the infusion rate. A survey was conducted to evaluate the performance of this system in a clinical setting. The infusion rate monitoring system was installed in 76 patients. Questionnaire responses were received from all 24 nurses involved in the study. Results regarding the usability and specifications of the infusion rate monitoring system showed that 24 (100%) were positive about the size, 23 (96%) about the weight and durability, and 22 (92%) about how the components were attached. The causes of alerts were 54 (35%) for infusion stoppage, 52 (34%) for overinfusion, and 32 (21%) for communication errors. The main causes of communication errors were transfers out of the patient room and transfers to the toilet in the patient room. This was due to the fact that the operating range of the system was 3 m from the receiver. To extend this range, new specifications need to be developed to cover a wider area.

Authors

Institutions

Publication Details

Journal
Hospital Administration and Medical Practices
Published
2024-12-30
DOI
https://doi.org/10.54844/hamp.2024.0057
Primary Topic
Intravenous Infusion Technology and Safety
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accident Prevention “Awareness” System for Intravenous Drip Infusion Operations - Construction and Evaluation

S Muraoka, Takao Orii, Hazuki Matsuo
Hospital Administration and Medical Practices
Intravenous Infusion Technology and Safety
article

Accident Prevention “Awareness” System for Intravenous Drip Infusion Operations - Construction and Evaluation

S Muraoka, Takao Orii, Hazuki Matsuo
article en

Abstract

The purpose of this study was to evaluate an infusion rate monitoring system (Tritech, Inc.) in a clinical setting and to identify challenges in using the system as an alternative to an infusion pump to prevent inappropriate infusion rates (excessive or insufficient). The system monitors the infusion rate and sends an alert to the nurse call system if the rate deviates from a preset limit. Unlike infusion pumps, this system does not adjust the infusion rate. A survey was conducted to evaluate the performance of this system in a clinical setting. The infusion rate monitoring system was installed in 76 patients. Questionnaire responses were received from all 24 nurses involved in the study. Results regarding the usability and specifications of the infusion rate monitoring system showed that 24 (100%) were positive about the size, 23 (96%) about the weight and durability, and 22 (92%) about how the components were attached. The causes of alerts were 54 (35%) for infusion stoppage, 52 (34%) for overinfusion, and 32 (21%) for communication errors. The main causes of communication errors were transfers out of the patient room and transfers to the toilet in the patient room. This was due to the fact that the operating range of the system was 3 m from the receiver. To extend this range, new specifications need to be developed to cover a wider area.

Hospital Administration and Medical PracticesVol. 3
NTT Medical Center (JP)
Good health and well-being
Openalex Percentile: Top 77%
Intravenous Infusion Technology and Safety
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.