Two-Dimensional Shear-Wave Elastography: Accuracy in Liver Fibrosis Staging Using Magnetic Resonance Elastography as the Reference Standard

<b>Background:</b> The accurate staging of liver fibrosis is crucial for managing chronic liver disease (CLD). Although magnetic resonance elastography (MRE) is the reference standard for noninvasive fibrosis assessment, its cost, specialized hardware, and operational demands restrict accessibility. In contrast, two-dimensional shear-wave elastography (2D-SWE) is more affordable, accessible, and widely integrated into routine ultrasound systems. <b>Objective:</b> Our aim was to determine the optimal 2D-SWE cut-offs for detecting significant fibrosis (≥F2) and evaluate its diagnostic performance across fibrosis stages. <b>Methods:</b> In this prospective study, 71 patients with suspected CLD underwent same-day MRE and 2D-SWE. MRE-defined cut-offs categorized fibrosis stages (≥3.5 kPa for significant fibrosis). Sensitivity, specificity, area under the receiver operating characteristic curve (AUROC), and likelihood ratios were calculated for various 2D-SWE thresholds. <b>Results:</b> At a 2D-SWE cut-off of 7.0 kPa, sensitivity for detecting ≥F2 fibrosis was 100% with a specificity of 85.7% and a positive likelihood ratio (LR+) of 7.0. Increasing the threshold to 8.0 kPa improved specificity to 91.8% while maintaining a sensitivity of 86.4% and achieving an AUROC of 0.89. For cirrhosis, a cut-off of 11.0 kPa achieved 100% sensitivity and 96.9% specificity. A 5.0 kPa cut-off reliably excluded abnormal stiffness with 89.1% sensitivity. <b>Conclusions:</b> Two-dimensional SWE is a reliable method for staging liver fibrosis. Thresholds of 7.0 kPa for screening significant fibrosis, 8.0 kPa for confirmation, and 11.0 kPa for diagnosing cirrhosis demonstrate high diagnostic accuracy. A 5.0 kPa cut-off effectively excludes abnormal liver stiffness.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2024-12-29
DOI
https://doi.org/10.3390/diagnostics15010062
Citations
5
Primary Topic
Liver Disease Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
1.93

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Two-Dimensional Shear-Wave Elastography: Accuracy in Liver Fibrosis Staging Using Magnetic Resonance Elastography as the Reference Standard

Puwitch Charoenchue, Apichat Tantraworasin, Taned Chitapanarux, Jiraporn Khorana et al.
5 citations
Diagnostics
Liver Disease Diagnosis and Treatment
1.93
article

Two-Dimensional Shear-Wave Elastography: Accuracy in Liver Fibrosis Staging Using Magnetic Resonance Elastography as the Reference Standard

Puwitch Charoenchue, Apichat Tantraworasin, Taned Chitapanarux, Jiraporn Khorana, Wittanee Na Chiangmai, Nakarin Inmutto, Suwalee Pojchamarnwiputh, Amonlaya Amantakul
article en
5 citations

Abstract

<b>Background:</b> The accurate staging of liver fibrosis is crucial for managing chronic liver disease (CLD). Although magnetic resonance elastography (MRE) is the reference standard for noninvasive fibrosis assessment, its cost, specialized hardware, and operational demands restrict accessibility. In contrast, two-dimensional shear-wave elastography (2D-SWE) is more affordable, accessible, and widely integrated into routine ultrasound systems. <b>Objective:</b> Our aim was to determine the optimal 2D-SWE cut-offs for detecting significant fibrosis (≥F2) and evaluate its diagnostic performance across fibrosis stages. <b>Methods:</b> In this prospective study, 71 patients with suspected CLD underwent same-day MRE and 2D-SWE. MRE-defined cut-offs categorized fibrosis stages (≥3.5 kPa for significant fibrosis). Sensitivity, specificity, area under the receiver operating characteristic curve (AUROC), and likelihood ratios were calculated for various 2D-SWE thresholds. <b>Results:</b> At a 2D-SWE cut-off of 7.0 kPa, sensitivity for detecting ≥F2 fibrosis was 100% with a specificity of 85.7% and a positive likelihood ratio (LR+) of 7.0. Increasing the threshold to 8.0 kPa improved specificity to 91.8% while maintaining a sensitivity of 86.4% and achieving an AUROC of 0.89. For cirrhosis, a cut-off of 11.0 kPa achieved 100% sensitivity and 96.9% specificity. A 5.0 kPa cut-off reliably excluded abnormal stiffness with 89.1% sensitivity. <b>Conclusions:</b> Two-dimensional SWE is a reliable method for staging liver fibrosis. Thresholds of 7.0 kPa for screening significant fibrosis, 8.0 kPa for confirmation, and 11.0 kPa for diagnosing cirrhosis demonstrate high diagnostic accuracy. A 5.0 kPa cut-off effectively excludes abnormal liver stiffness.

DiagnosticsVol. 15(1)
Chiang Mai University (TH)
Chiang Mai University, Faculty of Medicine, Chiang Mai University
Good health and well-being
Openalex Percentile: Top 12%
Liver Disease Diagnosis and Treatment
1.93
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.