Effects of different combinations of noise index and preset adaptive statistical iterative reconstruction Veo on the accuracy and image quality of bone mineral density measurements using fast kilovolt-switching dual-energy computed tomography: a phan
During the rapid kV-switching DECT scans, NI and ASiR-V, when combined with image quality and radiation dose, can improve the accuracy of BMD measurement, with the optimal parameter combination being NI =18 and ASiR-V<sub>60%</sub>.