Comparison of ZnS(Ag) Scintillator and Proportional Counter Tube for Alpha Detection in Thin-Layer Chromatography

(1) <b>Background</b>: Targeted alpha therapy is an emerging field in nuclear medicine driven by two advantages: overcoming resistance in cancer-suffering patients to beta therapies and the practical application of lower activities of <sup>212</sup>Pb- and <sup>225</sup>Ac-labelled peptides to achieve the same doses compared to beta therapy due to the highly cytotoxic nature of alpha particles. However, quality control of the <sup>212</sup>Pb/<sup>225</sup>Ac-radiopharmaceuticals remains a challenge due to the low activity levels used for therapy (100 kBq/kg) and the formation of several free daughter nuclides immediately after the formulation of patient doses; (2) <b>Methods</b>: The routine alpha detection on thin-layer chromatograms (TLC) of <sup>212</sup>Pb- and <sup>225</sup>Ac-labelled peptides using a MiniScanPRO+ scanner combined with an alpha detector head was compared with detection using an AR-2000 scanner equipped with an open proportional counter tube. Measurement time, resolution and validity were compared for both scanners; (3) <b>Results</b>: For <sup>225</sup>Ac, the quality control values of the radiochemical purity (RCP) were within the acceptance criteria 2 h after TLC development, regardless of when the TLC probe was taken. That is, if the TLC probe was taken 24 h after radiosynthesis, the true value of the RCP was not measured until 5 h after TLC development. For <sup>212</sup>Pb-labelled peptides, the probe sampling did not have a high impact on the value of the RCP for the MiniScanPRO+ and AR-2000. A difference was observed when measuring TLC with the AR-2000 in different modes; (4) <b>Conclusions</b>: The MiniScanPRO+ is fast, does not require additional equipment and can also measure the gamma spectrum, which may be important for some radiopharmaceutical production sites and regulatory authorities. The AR-2000 has a better signal-to-noise ratio, and this eliminates the need for additional waiting time after TLC development.

Authors

Institutions

Publication Details

Journal
Pharmaceuticals
Published
2024-12-28
DOI
https://doi.org/10.3390/ph18010026
Citations
2
Primary Topic
Radiopharmaceutical Chemistry and Applications
Type
article
Field-Weighted Citation Impact
0.81
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of ZnS(Ag) Scintillator and Proportional Counter Tube for Alpha Detection in Thin-Layer Chromatography

Ralph A. Bundschuh, Enrico Michler, Marc Pretze, Robert Freudenberg et al.
2 citations
Pharmaceuticals
Radiopharmaceutical Chemistry and Applications
0.81
article

Comparison of ZnS(Ag) Scintillator and Proportional Counter Tube for Alpha Detection in Thin-Layer Chromatography

Ralph A. Bundschuh, Enrico Michler, Marc Pretze, Robert Freudenberg, Holger Hartmann, Jörg Kotzerke, Jan Wendrich
article en
2 citations

Abstract

(1) <b>Background</b>: Targeted alpha therapy is an emerging field in nuclear medicine driven by two advantages: overcoming resistance in cancer-suffering patients to beta therapies and the practical application of lower activities of <sup>212</sup>Pb- and <sup>225</sup>Ac-labelled peptides to achieve the same doses compared to beta therapy due to the highly cytotoxic nature of alpha particles. However, quality control of the <sup>212</sup>Pb/<sup>225</sup>Ac-radiopharmaceuticals remains a challenge due to the low activity levels used for therapy (100 kBq/kg) and the formation of several free daughter nuclides immediately after the formulation of patient doses; (2) <b>Methods</b>: The routine alpha detection on thin-layer chromatograms (TLC) of <sup>212</sup>Pb- and <sup>225</sup>Ac-labelled peptides using a MiniScanPRO+ scanner combined with an alpha detector head was compared with detection using an AR-2000 scanner equipped with an open proportional counter tube. Measurement time, resolution and validity were compared for both scanners; (3) <b>Results</b>: For <sup>225</sup>Ac, the quality control values of the radiochemical purity (RCP) were within the acceptance criteria 2 h after TLC development, regardless of when the TLC probe was taken. That is, if the TLC probe was taken 24 h after radiosynthesis, the true value of the RCP was not measured until 5 h after TLC development. For <sup>212</sup>Pb-labelled peptides, the probe sampling did not have a high impact on the value of the RCP for the MiniScanPRO+ and AR-2000. A difference was observed when measuring TLC with the AR-2000 in different modes; (4) <b>Conclusions</b>: The MiniScanPRO+ is fast, does not require additional equipment and can also measure the gamma spectrum, which may be important for some radiopharmaceutical production sites and regulatory authorities. The AR-2000 has a better signal-to-noise ratio, and this eliminates the need for additional waiting time after TLC development.

PharmaceuticalsVol. 18(1)
University Hospital Carl Gustav Carus (DE)
Good health and well-being
Openalex Percentile: Top 23%
Radiopharmaceutical Chemistry and Applications
0.81
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.