Analyzing and Forecasting Laboratory Energy Consumption Patterns Using Autoregressive Integrated Moving Average Models

This study applied ARIMA modeling to analyze the energy consumption patterns of laboratory equipment over one month, focusing on enhancing energy management in the laboratory. By explicitly examining AC and DC equipment, this study obtained detailed daily operating cycles and periods of inactivity. Advanced differencing and diagnostic checks were used to verify model accuracy and white noise characteristics through enhanced Dickey–Fuller testing and residual analysis. The results demonstrate the model’s accuracy in predicting energy consumption, providing valuable insights into the use of the model. This study highlights the adaptability and validity of the ARIMA model in laboratory environments, contributing to more competent laboratory energy management practices.

Authors

Institutions

Publication Details

Journal
Laboratories
Published
2024-12-30
DOI
https://doi.org/10.3390/laboratories2010002
Citations
9
Primary Topic
Building Energy and Comfort Optimization
Type
article
Field-Weighted Citation Impact
2.39

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Analyzing and Forecasting Laboratory Energy Consumption Patterns Using Autoregressive Integrated Moving Average Models

Chee Keong Lee, Yitong Niu, Haoran Jiang, Cheu Peng Leh et al.
9 citations
Laboratories
Building Energy and Comfort Optimization
2.39
article

Analyzing and Forecasting Laboratory Energy Consumption Patterns Using Autoregressive Integrated Moving Average Models

Chee Keong Lee, Yitong Niu, Haoran Jiang, Cheu Peng Leh, Xiongjie Jia
article en
9 citations

Abstract

This study applied ARIMA modeling to analyze the energy consumption patterns of laboratory equipment over one month, focusing on enhancing energy management in the laboratory. By explicitly examining AC and DC equipment, this study obtained detailed daily operating cycles and periods of inactivity. Advanced differencing and diagnostic checks were used to verify model accuracy and white noise characteristics through enhanced Dickey–Fuller testing and residual analysis. The results demonstrate the model’s accuracy in predicting energy consumption, providing valuable insights into the use of the model. This study highlights the adaptability and validity of the ARIMA model in laboratory environments, contributing to more competent laboratory energy management practices.

LaboratoriesVol. 2(1)
Universiti Sains Malaysia (MY), Tianjin University (CN)
Ministry of Higher Education, Malaysia
Affordable and clean energy
Openalex Percentile: Top 12%
Building Energy and Comfort Optimization
2.39
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.