A Comparative Analysis of the Effects of Misaligning Different Trifocal Intraocular Lenses
<b>Objectives:</b> This laboratory study aims to assess the effects of misaligning different trifocal intraocular lenses (IOLs) under varying spectral and corneal spherical aberration (SA) conditions. <b>Methods:</b> With an IOL metrology device under monochromatic and polychromatic conditions, the following models were studied: AT ELANA 841P, AT LISA Tri 839MP, FineVision HP POD F, Acrysof IQ PanOptix, and Tecnis Synergy ZFR00V. The SA was simulated using an aberration-free and average-SA cornea. The modulation transfer function (MTF) was measured at different pupil sizes for the on- and off-axis lens positions. <b>Results:</b> The IOLs exhibited varying responses to decentration up to 1 mm, showing the lowest impact in polychromatic light. The least affected was AT ELANA, with an MTF loss of 15.7% to 28.4% at 50 lp/mm across the studied conditions. It was followed by PanOptix and FineVision, with the MTF loss ranging from 19.1% to 36.0% and from 21.2% to 46.6%. AT LISA showed a more substantial reduction, i.e., 41.2% to 64.8%, but it was still lower than that of Synergy (51.1% to 78.8%). When decentration was induced at a 4.5 mm distance, its effect was more evident in conditions that were closer to each IOL's SA correction. A tilt of 5° had a lesser impact than 1 mm decentration, with the effect being more severe at 4.5 mm. <b>Conclusions:</b> The off-axis position affects the optical quality of trifocal IOLs. Low- rather than high-SA-correcting trifocals perform better under misalignment. In polychromatic light, the impact of misalignment is less evident, suggesting a potential mitigating effect of chromatic aberration.
Authors
- Gerd Uwe Auffarth (ORCID: https://orcid.org/0000-0002-6927-5251)
- Ramin Khoramnia (ORCID: https://orcid.org/0000-0002-6237-7773)
- Weijia Yan (ORCID: https://orcid.org/0000-0002-3069-3333)
- Grzegorz Łabuz
Institutions
- Heidelberg University (DE)
- University Hospital Heidelberg (DE)
- Second Affiliated Hospital of Zhejiang University (CN)
Publication Details
- Journal
- Journal of Clinical Medicine
- Published
- 2024-12-31
- DOI
- https://doi.org/10.3390/jcm14010187
- Citations
- 6
- Primary Topic
- Ophthalmology and Visual Impairment Studies
- Type
- article
- Field-Weighted Citation Impact
- 2.32
Funders
- Carl Zeiss Meditec AG
- Klaus Tschira Stiftung