A Comparative Analysis of the Effects of Misaligning Different Trifocal Intraocular Lenses

<b>Objectives:</b> This laboratory study aims to assess the effects of misaligning different trifocal intraocular lenses (IOLs) under varying spectral and corneal spherical aberration (SA) conditions. <b>Methods:</b> With an IOL metrology device under monochromatic and polychromatic conditions, the following models were studied: AT ELANA 841P, AT LISA Tri 839MP, FineVision HP POD F, Acrysof IQ PanOptix, and Tecnis Synergy ZFR00V. The SA was simulated using an aberration-free and average-SA cornea. The modulation transfer function (MTF) was measured at different pupil sizes for the on- and off-axis lens positions. <b>Results:</b> The IOLs exhibited varying responses to decentration up to 1 mm, showing the lowest impact in polychromatic light. The least affected was AT ELANA, with an MTF loss of 15.7% to 28.4% at 50 lp/mm across the studied conditions. It was followed by PanOptix and FineVision, with the MTF loss ranging from 19.1% to 36.0% and from 21.2% to 46.6%. AT LISA showed a more substantial reduction, i.e., 41.2% to 64.8%, but it was still lower than that of Synergy (51.1% to 78.8%). When decentration was induced at a 4.5 mm distance, its effect was more evident in conditions that were closer to each IOL's SA correction. A tilt of 5° had a lesser impact than 1 mm decentration, with the effect being more severe at 4.5 mm. <b>Conclusions:</b> The off-axis position affects the optical quality of trifocal IOLs. Low- rather than high-SA-correcting trifocals perform better under misalignment. In polychromatic light, the impact of misalignment is less evident, suggesting a potential mitigating effect of chromatic aberration.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2024-12-31
DOI
https://doi.org/10.3390/jcm14010187
Citations
6
Primary Topic
Ophthalmology and Visual Impairment Studies
Type
article
Field-Weighted Citation Impact
2.32

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Comparative Analysis of the Effects of Misaligning Different Trifocal Intraocular Lenses

Gerd Uwe Auffarth, Ramin Khoramnia, Weijia Yan, Grzegorz Łabuz
6 citations
Journal of Clinical Medicine
Ophthalmology and Visual Impairment Studies
2.32
article

A Comparative Analysis of the Effects of Misaligning Different Trifocal Intraocular Lenses

Gerd Uwe Auffarth, Ramin Khoramnia, Weijia Yan, Grzegorz Łabuz
article en
6 citations

Abstract

<b>Objectives:</b> This laboratory study aims to assess the effects of misaligning different trifocal intraocular lenses (IOLs) under varying spectral and corneal spherical aberration (SA) conditions. <b>Methods:</b> With an IOL metrology device under monochromatic and polychromatic conditions, the following models were studied: AT ELANA 841P, AT LISA Tri 839MP, FineVision HP POD F, Acrysof IQ PanOptix, and Tecnis Synergy ZFR00V. The SA was simulated using an aberration-free and average-SA cornea. The modulation transfer function (MTF) was measured at different pupil sizes for the on- and off-axis lens positions. <b>Results:</b> The IOLs exhibited varying responses to decentration up to 1 mm, showing the lowest impact in polychromatic light. The least affected was AT ELANA, with an MTF loss of 15.7% to 28.4% at 50 lp/mm across the studied conditions. It was followed by PanOptix and FineVision, with the MTF loss ranging from 19.1% to 36.0% and from 21.2% to 46.6%. AT LISA showed a more substantial reduction, i.e., 41.2% to 64.8%, but it was still lower than that of Synergy (51.1% to 78.8%). When decentration was induced at a 4.5 mm distance, its effect was more evident in conditions that were closer to each IOL's SA correction. A tilt of 5° had a lesser impact than 1 mm decentration, with the effect being more severe at 4.5 mm. <b>Conclusions:</b> The off-axis position affects the optical quality of trifocal IOLs. Low- rather than high-SA-correcting trifocals perform better under misalignment. In polychromatic light, the impact of misalignment is less evident, suggesting a potential mitigating effect of chromatic aberration.

Journal of Clinical MedicineVol. 14(1)
Heidelberg University (DE), University Hospital Heidelberg (DE), Second Affiliated Hospital of Zhejiang University (CN)
Carl Zeiss Meditec AG, Klaus Tschira Stiftung
Openalex Percentile: Top 10%
Ophthalmology and Visual Impairment Studies
2.32
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.