Detailed improved uncertainty analysis for micropipette calibration in the range below 10 μL

Micropipettes are the most efficient tool for performing volume measurements in fields such as medicine, chemistry, biology, pharmacy, and genetics. Laboratories must ensure that the data generated by this equipment are accurate in order to compute uncertainties and calibrate micropipettes appropriately. This work aims to define the optimal method for micropipette calibration down to 1 μL under the given measurement conditions, with error sources to be identified, investigated, and fixed. A calculation model for measurement uncertainty is also provided, which is a core and main part of this paper. Detailed uncertainty analysis, including models and related derivations are also presented. At the calibration level of 1 μL, an uncertainty level of 0.019 μL ( k = 2) could be attained.

Authors

Institutions

Publication Details

Journal
Measurement Sensors
Published
2024-12-31
DOI
https://doi.org/10.1016/j.measen.2024.101528
Primary Topic
Integrated Circuits and Semiconductor Failure Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Detailed improved uncertainty analysis for micropipette calibration in the range below 10 μL

Khaled M. Ahmed, Abdulkarim A. AlShahrany
Measurement Sensors
Integrated Circuits and Semiconductor Failure Analysis
article

Detailed improved uncertainty analysis for micropipette calibration in the range below 10 μL

Khaled M. Ahmed, Abdulkarim A. AlShahrany
article en

Abstract

Micropipettes are the most efficient tool for performing volume measurements in fields such as medicine, chemistry, biology, pharmacy, and genetics. Laboratories must ensure that the data generated by this equipment are accurate in order to compute uncertainties and calibrate micropipettes appropriately. This work aims to define the optimal method for micropipette calibration down to 1 μL under the given measurement conditions, with error sources to be identified, investigated, and fixed. A calculation model for measurement uncertainty is also provided, which is a core and main part of this paper. Detailed uncertainty analysis, including models and related derivations are also presented. At the calibration level of 1 μL, an uncertainty level of 0.019 μL ( k = 2) could be attained.

Measurement SensorsVol. 38
National Institute of Standards (EG), National Measurement Laboratory (AU)
Openalex Percentile: Top 77%
Integrated Circuits and Semiconductor Failure Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.