Interface Pressures Derived from a Calibrated Bandage Applied for Compression Therapy

<b>Objectives:</b> We sought to clarify the interface pressure (IP) and its variation by applying Biflex16, a calibrated bandage, to the lower leg. <b>Methods:</b> In Study I, 50 participants applied a bandage to the lower leg of a single subject in two ways: first, with 50% overlap, while the calibration rectangle became a square (Application 1), and then with 50% overlap without intentional stretch (Application 2) which served as a control. In Study II, another 51 participants applied the bandage to their lower leg via Application 1. The IP was measured at the level of the transposition of the medial gastrocnemius muscle into the Achilles tendon (B1). <b>Results:</b> In Study I, the median IP (37 mmHg) and interquartile range (IQR; 9 mmHg) in the standing position were the same for Applications 1 and 2. In Study II, the obtained IP and IQR values were 38 and 12 mmHg, respectively, in the sitting position. This IP was similar to that obtained in Study I, and no correlation was found between IP and leg circumference. <b>Conclusions:</b> The variation in the IP obtained by the calibrated bandage was reasonably small when applied via Application 1. The obtained IPs did not correlate with the leg circumference.

Authors

Institutions

Publication Details

Journal
Annals of Vascular Diseases
Published
2024-12-31
DOI
https://doi.org/10.3400/avd.oa.24-00103
Primary Topic
Diagnosis and Treatment of Venous Diseases
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Interface Pressures Derived from a Calibrated Bandage Applied for Compression Therapy

Ryo Suzuki, Kotaro Suehiro, Yuriko Takeuchi, Ryunosuke Sakamoto et al.
Annals of Vascular Diseases
Diagnosis and Treatment of Venous Diseases
article

Interface Pressures Derived from a Calibrated Bandage Applied for Compression Therapy

Ryo Suzuki, Kotaro Suehiro, Yuriko Takeuchi, Ryunosuke Sakamoto, Hitoshi Sakuda, Hiroshi Kurazumi, Takahiro Mizoguchi, Kimikazu Hamano, Takasuke Harada
article en

Abstract

<b>Objectives:</b> We sought to clarify the interface pressure (IP) and its variation by applying Biflex16, a calibrated bandage, to the lower leg. <b>Methods:</b> In Study I, 50 participants applied a bandage to the lower leg of a single subject in two ways: first, with 50% overlap, while the calibration rectangle became a square (Application 1), and then with 50% overlap without intentional stretch (Application 2) which served as a control. In Study II, another 51 participants applied the bandage to their lower leg via Application 1. The IP was measured at the level of the transposition of the medial gastrocnemius muscle into the Achilles tendon (B1). <b>Results:</b> In Study I, the median IP (37 mmHg) and interquartile range (IQR; 9 mmHg) in the standing position were the same for Applications 1 and 2. In Study II, the obtained IP and IQR values were 38 and 12 mmHg, respectively, in the sitting position. This IP was similar to that obtained in Study I, and no correlation was found between IP and leg circumference. <b>Conclusions:</b> The variation in the IP obtained by the calibrated bandage was reasonably small when applied via Application 1. The obtained IPs did not correlate with the leg circumference.

Annals of Vascular DiseasesVol. 18(1)
Yamaguchi University (JP), Japanese Society for Dialysis Therapy (JP)
Affordable and clean energy
Openalex Percentile: Top 55%
Diagnosis and Treatment of Venous Diseases
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Interface Pressures Derived from a Calibrated Bandage Applied for Compression Therapy — Ryo Suzuki, Kotaro Suehiro, et al. · Annals of Vascular Diseases (2024) | TGRS Research Map | TGRS